Side viewing micro-endoscope optical system with acromatized objective lens

High resolution microendoscopy (HRME) is a novel optical technique which utilizes optical system to view directly sub-cellular structures. Major obstacles of HRME which uses forward view optical system in achieving high quality image of sub cellular structures in GI tract wall are short depth of fie...

Full description

Bibliographic Details
Main Author: Doraj-Raj, Nanthakumar
Format: Thesis
Language:English
Published: 2014
Subjects:
Online Access:http://psasir.upm.edu.my/id/eprint/55787/
http://psasir.upm.edu.my/id/eprint/55787/1/ITMA%202014%204RR.pdf
_version_ 1848852899925327872
author Doraj-Raj, Nanthakumar
author_facet Doraj-Raj, Nanthakumar
author_sort Doraj-Raj, Nanthakumar
building UPM Institutional Repository
collection Online Access
description High resolution microendoscopy (HRME) is a novel optical technique which utilizes optical system to view directly sub-cellular structures. Major obstacles of HRME which uses forward view optical system in achieving high quality image of sub cellular structures in GI tract wall are short depth of field and small field of view. Vibration or movement of the optical system during probing resulted in low image acquisition quality. In this thesis, design and optimization of side view micro endoscope with magnification of 3X optical system using fold mirror,doublets and aspheric lens are presented. Fold mirror incorporated into optical system with a purpose of bending light path which allows GI tract wall viewed side way while rotating for circumferential scanning. In order to minimize aberrations that degrade images created by fold mirror, aspheric lens with conic shape added before fold mirror which functions to collimate rays before reaching fold mirror. Next relationship between working distance, numerical aperture of object and image space and airy disc established to determine magnification and resolution of the optical system. Then 4 doublets lens with varying dispersion ratio used in order to achieve achromatization which functions to increase field of view, working wavelength range, minimizes aberrations and achieved acceptable image qualities. Optimization of optical system performed for optimum performance achievements. Tolerance analysis simulation performed by adding components tolerances in predicting manufacturability of the conceptual design. Simulation results across field of view for wavelength range from 0.53 μm to 0.63 μm presented. Optimum MTF value above 20% recorded for spatial frequency of 240 line/mm at axis. MTF value above 20% recorded for spatial frequency of 170 line/mm at the edge. Maximum drop of 20% of MTF achieved after adding components tolerance limit. This technology can enhance image acquisition quality of HRME while viewing sub cellular tissues such as cytoplasm and nucleus as side view allows circumferential scanning of optical system which simplified movement mechanism. The optical system able to fit into capsule that works as pressure plate to maintain stability. In future developing prototype of the optical system can be used a basis for developing automated circumferential scanning of GI tract.
first_indexed 2025-11-15T10:45:25Z
format Thesis
id upm-55787
institution Universiti Putra Malaysia
institution_category Local University
language English
last_indexed 2025-11-15T10:45:25Z
publishDate 2014
recordtype eprints
repository_type Digital Repository
spelling upm-557872017-06-30T05:13:38Z http://psasir.upm.edu.my/id/eprint/55787/ Side viewing micro-endoscope optical system with acromatized objective lens Doraj-Raj, Nanthakumar High resolution microendoscopy (HRME) is a novel optical technique which utilizes optical system to view directly sub-cellular structures. Major obstacles of HRME which uses forward view optical system in achieving high quality image of sub cellular structures in GI tract wall are short depth of field and small field of view. Vibration or movement of the optical system during probing resulted in low image acquisition quality. In this thesis, design and optimization of side view micro endoscope with magnification of 3X optical system using fold mirror,doublets and aspheric lens are presented. Fold mirror incorporated into optical system with a purpose of bending light path which allows GI tract wall viewed side way while rotating for circumferential scanning. In order to minimize aberrations that degrade images created by fold mirror, aspheric lens with conic shape added before fold mirror which functions to collimate rays before reaching fold mirror. Next relationship between working distance, numerical aperture of object and image space and airy disc established to determine magnification and resolution of the optical system. Then 4 doublets lens with varying dispersion ratio used in order to achieve achromatization which functions to increase field of view, working wavelength range, minimizes aberrations and achieved acceptable image qualities. Optimization of optical system performed for optimum performance achievements. Tolerance analysis simulation performed by adding components tolerances in predicting manufacturability of the conceptual design. Simulation results across field of view for wavelength range from 0.53 μm to 0.63 μm presented. Optimum MTF value above 20% recorded for spatial frequency of 240 line/mm at axis. MTF value above 20% recorded for spatial frequency of 170 line/mm at the edge. Maximum drop of 20% of MTF achieved after adding components tolerance limit. This technology can enhance image acquisition quality of HRME while viewing sub cellular tissues such as cytoplasm and nucleus as side view allows circumferential scanning of optical system which simplified movement mechanism. The optical system able to fit into capsule that works as pressure plate to maintain stability. In future developing prototype of the optical system can be used a basis for developing automated circumferential scanning of GI tract. 2014-10 Thesis NonPeerReviewed application/pdf en http://psasir.upm.edu.my/id/eprint/55787/1/ITMA%202014%204RR.pdf Doraj-Raj, Nanthakumar (2014) Side viewing micro-endoscope optical system with acromatized objective lens. Masters thesis, Universiti Putra Malaysia. Lenses - Design and construction Geometrical optics Diagnostic imaging
spellingShingle Lenses - Design and construction
Geometrical optics
Diagnostic imaging
Doraj-Raj, Nanthakumar
Side viewing micro-endoscope optical system with acromatized objective lens
title Side viewing micro-endoscope optical system with acromatized objective lens
title_full Side viewing micro-endoscope optical system with acromatized objective lens
title_fullStr Side viewing micro-endoscope optical system with acromatized objective lens
title_full_unstemmed Side viewing micro-endoscope optical system with acromatized objective lens
title_short Side viewing micro-endoscope optical system with acromatized objective lens
title_sort side viewing micro-endoscope optical system with acromatized objective lens
topic Lenses - Design and construction
Geometrical optics
Diagnostic imaging
url http://psasir.upm.edu.my/id/eprint/55787/
http://psasir.upm.edu.my/id/eprint/55787/1/ITMA%202014%204RR.pdf