The validity and reliability of motion analysis in patellar tendon reflex assessment

BACKGROUND: The deep tendon reflex assessments that are essential to the accurate diagnosis of neurological or neuromuscular disorders are conducted subjectively in clinical neurology. Our aim was to assess deep tendon reflexes objectively with a new reflex quantification method. METHODOLOGY/PRINCIP...

Full description

Bibliographic Details
Main Authors: Tham, L.K., Abu Osman, N.A., Wan Abas, W.A.B., Lim, K.S.
Format: Article
Published: Public Library of Science 2013
Subjects:
Online Access:http://eprints.um.edu.my/7101/
id um-7101
recordtype eprints
spelling um-71012018-08-09T09:08:45Z The validity and reliability of motion analysis in patellar tendon reflex assessment Tham, L.K. Abu Osman, N.A. Wan Abas, W.A.B. Lim, K.S. TS Manufactures BACKGROUND: The deep tendon reflex assessments that are essential to the accurate diagnosis of neurological or neuromuscular disorders are conducted subjectively in clinical neurology. Our aim was to assess deep tendon reflexes objectively with a new reflex quantification method. METHODOLOGY/PRINCIPAL FINDINGS: The present study used a motion analysis technique to collect quantitative measurements for both the input and output of normal patellar tendon reflex. Reflex responses were measured as knee angles. The patellar tendon reflexes of 100 healthy subjects were examined using 6 levels of tendon taps, where all the assessments were captured using motion capture system. A linear relationship was found between the experimental maximum tapping velocity and tapping angle (coefficient of determination = 0.989), which was consistent with the theoretical values. Tapping velocities were predictable according to tapping angles. The findings proved the reproducibility of tapping method in producing consistent input. The reflex amplitude was consistent between two randomly assigned groups, and linearly proportionate to the tapping velocity. CONCLUSIONS/SIGNIFICANCE: The findings on reflex amplitude indicate that motion analysis is a valid and reliable method of assessing and measuring deep tendon reflexes. Public Library of Science 2013 Article PeerReviewed Tham, L.K.; Abu Osman, N.A.; Wan Abas, W.A.B.; Lim, K.S. (2013) The validity and reliability of motion analysis in patellar tendon reflex assessment. PLoS ONE <http://eprints.um.edu.my/view/publication/PLoS_ONE.html>, 8 (2). e55702. ISSN 1932-6203 http://eprints.um.edu.my/7101/
repository_type Digital Repository
institution_category Local University
institution University Malaya
building UM Research Repository
collection Online Access
topic TS Manufactures
spellingShingle TS Manufactures
Tham, L.K.
Abu Osman, N.A.
Wan Abas, W.A.B.
Lim, K.S.
The validity and reliability of motion analysis in patellar tendon reflex assessment
description BACKGROUND: The deep tendon reflex assessments that are essential to the accurate diagnosis of neurological or neuromuscular disorders are conducted subjectively in clinical neurology. Our aim was to assess deep tendon reflexes objectively with a new reflex quantification method. METHODOLOGY/PRINCIPAL FINDINGS: The present study used a motion analysis technique to collect quantitative measurements for both the input and output of normal patellar tendon reflex. Reflex responses were measured as knee angles. The patellar tendon reflexes of 100 healthy subjects were examined using 6 levels of tendon taps, where all the assessments were captured using motion capture system. A linear relationship was found between the experimental maximum tapping velocity and tapping angle (coefficient of determination = 0.989), which was consistent with the theoretical values. Tapping velocities were predictable according to tapping angles. The findings proved the reproducibility of tapping method in producing consistent input. The reflex amplitude was consistent between two randomly assigned groups, and linearly proportionate to the tapping velocity. CONCLUSIONS/SIGNIFICANCE: The findings on reflex amplitude indicate that motion analysis is a valid and reliable method of assessing and measuring deep tendon reflexes.
format Article
author Tham, L.K.
Abu Osman, N.A.
Wan Abas, W.A.B.
Lim, K.S.
author_facet Tham, L.K.
Abu Osman, N.A.
Wan Abas, W.A.B.
Lim, K.S.
author_sort Tham, L.K.
title The validity and reliability of motion analysis in patellar tendon reflex assessment
title_short The validity and reliability of motion analysis in patellar tendon reflex assessment
title_full The validity and reliability of motion analysis in patellar tendon reflex assessment
title_fullStr The validity and reliability of motion analysis in patellar tendon reflex assessment
title_full_unstemmed The validity and reliability of motion analysis in patellar tendon reflex assessment
title_sort validity and reliability of motion analysis in patellar tendon reflex assessment
publisher Public Library of Science
publishDate 2013
url http://eprints.um.edu.my/7101/
first_indexed 2018-09-06T05:24:06Z
last_indexed 2018-09-06T05:24:06Z
_version_ 1610834454144614400