Stability of Standard Electrolytic Conductivity Solutions in Glass Containers
The stability of solutions having an electrolytic conductivity, κ, of 5 μS/cm to 100 000 μS/cm packaged in glass screw-cap bottles, glass serum bottles, and glass ampoules was monitored for 1 year to 2 years. The conductivity was determined by measuring the ac resistance of the solution. Mass loss w...
Main Author: | |
---|---|
Format: | Online |
Language: | English |
Published: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2002
|
Online Access: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4861383/ |