EBT2 film dosimetry using Imaged software analysis in comparison to Verisoft software at High Energy Electron Beam
This study was conducted, to compare ImageJ and Verisoft softwares in EBT2 Gafchromic film dosimetry in High Energy Electron Beam. Film was cut into 3 cm x 3 cm dimension. 6 MeV and 9MeV energy use under Siemens Primus (LINAC), for calibration purpose, the film was exposed to dose 0 cGy to 600 cGy...
| Main Author: | |
|---|---|
| Format: | Monograph |
| Language: | English |
| Published: |
Universiti Sains Malaysia
2016
|
| Subjects: | |
| Online Access: | http://eprints.usm.my/62721/ http://eprints.usm.my/62721/1/MOHD%20FAHRIZUAN%20BIN%20MD%20YUSOP%20-%20e.pdf |
| _version_ | 1848885070604009472 |
|---|---|
| author | Yusop, Mohd Fahrizuan Md |
| author_facet | Yusop, Mohd Fahrizuan Md |
| author_sort | Yusop, Mohd Fahrizuan Md |
| building | USM Institutional Repository |
| collection | Online Access |
| description | This study was conducted, to compare ImageJ and Verisoft softwares in EBT2
Gafchromic film dosimetry in High Energy Electron Beam. Film was cut into 3 cm x 3 cm dimension. 6 MeV and 9MeV energy use under Siemens Primus (LINAC), for calibration purpose, the film was exposed to dose 0 cGy to 600 cGy and for measuring
FDD and beam profile, another film was irradiated with 400MU monitor unit. The
irradiated films were scanned with Epson Expression 10 000XL flatbed scanner. The
percentage depth dose (PDD) and beam profile of 6 MeV and 9 MeV electron beams were analyzed by using ImageJ and Verisoft software. Result of this study shows, the discrepancy (% of error) of ImageJ towards Verisoft for all depth are 4.05 % and
discrepancy (% of error) comparison between ImageJ software with Ionization
Chamber reading are 29.4 %. The highest beam flatness of Verisoft and ImageJ
software are 5.88 % and 8 % respectively, the highest penumbra size for Verisoft and
ImageJ software are 4 cm both. In conclusions, ImageJ software can be use as
software in electron beam dosimetric analysis since it has less percent of error
compare to Verisoft software and Ionization Chamber reading. |
| first_indexed | 2025-11-15T19:16:45Z |
| format | Monograph |
| id | usm-62721 |
| institution | Universiti Sains Malaysia |
| institution_category | Local University |
| language | English |
| last_indexed | 2025-11-15T19:16:45Z |
| publishDate | 2016 |
| publisher | Universiti Sains Malaysia |
| recordtype | eprints |
| repository_type | Digital Repository |
| spelling | usm-627212025-09-07T08:08:15Z http://eprints.usm.my/62721/ EBT2 film dosimetry using Imaged software analysis in comparison to Verisoft software at High Energy Electron Beam Yusop, Mohd Fahrizuan Md R Medicine (General) RA Public aspects of medicine This study was conducted, to compare ImageJ and Verisoft softwares in EBT2 Gafchromic film dosimetry in High Energy Electron Beam. Film was cut into 3 cm x 3 cm dimension. 6 MeV and 9MeV energy use under Siemens Primus (LINAC), for calibration purpose, the film was exposed to dose 0 cGy to 600 cGy and for measuring FDD and beam profile, another film was irradiated with 400MU monitor unit. The irradiated films were scanned with Epson Expression 10 000XL flatbed scanner. The percentage depth dose (PDD) and beam profile of 6 MeV and 9 MeV electron beams were analyzed by using ImageJ and Verisoft software. Result of this study shows, the discrepancy (% of error) of ImageJ towards Verisoft for all depth are 4.05 % and discrepancy (% of error) comparison between ImageJ software with Ionization Chamber reading are 29.4 %. The highest beam flatness of Verisoft and ImageJ software are 5.88 % and 8 % respectively, the highest penumbra size for Verisoft and ImageJ software are 4 cm both. In conclusions, ImageJ software can be use as software in electron beam dosimetric analysis since it has less percent of error compare to Verisoft software and Ionization Chamber reading. Universiti Sains Malaysia 2016 Monograph NonPeerReviewed application/pdf en http://eprints.usm.my/62721/1/MOHD%20FAHRIZUAN%20BIN%20MD%20YUSOP%20-%20e.pdf Yusop, Mohd Fahrizuan Md (2016) EBT2 film dosimetry using Imaged software analysis in comparison to Verisoft software at High Energy Electron Beam. Project Report. Universiti Sains Malaysia. (Submitted) |
| spellingShingle | R Medicine (General) RA Public aspects of medicine Yusop, Mohd Fahrizuan Md EBT2 film dosimetry using Imaged software analysis in comparison to Verisoft software at High Energy Electron Beam |
| title | EBT2 film dosimetry using Imaged software analysis in
comparison to Verisoft software at High Energy Electron Beam |
| title_full | EBT2 film dosimetry using Imaged software analysis in
comparison to Verisoft software at High Energy Electron Beam |
| title_fullStr | EBT2 film dosimetry using Imaged software analysis in
comparison to Verisoft software at High Energy Electron Beam |
| title_full_unstemmed | EBT2 film dosimetry using Imaged software analysis in
comparison to Verisoft software at High Energy Electron Beam |
| title_short | EBT2 film dosimetry using Imaged software analysis in
comparison to Verisoft software at High Energy Electron Beam |
| title_sort | ebt2 film dosimetry using imaged software analysis in
comparison to verisoft software at high energy electron beam |
| topic | R Medicine (General) RA Public aspects of medicine |
| url | http://eprints.usm.my/62721/ http://eprints.usm.my/62721/1/MOHD%20FAHRIZUAN%20BIN%20MD%20YUSOP%20-%20e.pdf |