Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit

Numerical studies of blood flow system of aorta coronary sinus conduit were carried out using ANSYS™ CFD simulation. Pressure drop was investigated to achieve from 80 mmHg to 15 mmHg with a 3D model blood conduit. From the analysis of coronary sinus conduit, it is found that a narrow tube needs to b...

Full description

Bibliographic Details
Main Authors: Hussain, Siti Aslina, Tan, Hong Tat, Hamid, Mohd Ismail, Abdullah, Norhafizah, Idris, Azni
Format: Conference or Workshop Item
Language:English
English
Published: 2010
Subjects:
Online Access:http://psasir.upm.edu.my/id/eprint/9402/
http://psasir.upm.edu.my/id/eprint/9402/1/ID%209402.pdf
_version_ 1848841136663166976
author Hussain, Siti Aslina
Tan, Hong Tat
Hamid, Mohd Ismail
Abdullah, Norhafizah
Idris, Azni
author_facet Hussain, Siti Aslina
Tan, Hong Tat
Hamid, Mohd Ismail
Abdullah, Norhafizah
Idris, Azni
author_sort Hussain, Siti Aslina
building UPM Institutional Repository
collection Online Access
description Numerical studies of blood flow system of aorta coronary sinus conduit were carried out using ANSYS™ CFD simulation. Pressure drop was investigated to achieve from 80 mmHg to 15 mmHg with a 3D model blood conduit. From the analysis of coronary sinus conduit, it is found that a narrow tube needs to be incorporated into the conduit produced. This is to induce a venturi effect to reduce the pressure of blood within a specific throat length. As a conclusion, a model of 3 mm inlet and throat diameter of 1.13 mm show satisfactory result for pressure reduction from 80 mmHg to 15 mmHg.
first_indexed 2025-11-15T07:38:27Z
format Conference or Workshop Item
id upm-9402
institution Universiti Putra Malaysia
institution_category Local University
language English
English
last_indexed 2025-11-15T07:38:27Z
publishDate 2010
recordtype eprints
repository_type Digital Repository
spelling upm-94022014-02-10T03:36:22Z http://psasir.upm.edu.my/id/eprint/9402/ Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit Hussain, Siti Aslina Tan, Hong Tat Hamid, Mohd Ismail Abdullah, Norhafizah Idris, Azni Numerical studies of blood flow system of aorta coronary sinus conduit were carried out using ANSYS™ CFD simulation. Pressure drop was investigated to achieve from 80 mmHg to 15 mmHg with a 3D model blood conduit. From the analysis of coronary sinus conduit, it is found that a narrow tube needs to be incorporated into the conduit produced. This is to induce a venturi effect to reduce the pressure of blood within a specific throat length. As a conclusion, a model of 3 mm inlet and throat diameter of 1.13 mm show satisfactory result for pressure reduction from 80 mmHg to 15 mmHg. 2010-07-05 Conference or Workshop Item NonPeerReviewed application/pdf en http://psasir.upm.edu.my/id/eprint/9402/1/ID%209402.pdf Hussain, Siti Aslina and Tan, Hong Tat and Hamid, Mohd Ismail and Abdullah, Norhafizah and Idris, Azni (2010) Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit. In: 6th International Conference on Diffusion in Solids and Liquids – DSL-2010, 5 - 7 July, 2010, Paris, France. . (Unpublished) Algebra - Graphic methods Aorta Sinusitis English
spellingShingle Algebra - Graphic methods
Aorta
Sinusitis
Hussain, Siti Aslina
Tan, Hong Tat
Hamid, Mohd Ismail
Abdullah, Norhafizah
Idris, Azni
Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit
title Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit
title_full Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit
title_fullStr Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit
title_full_unstemmed Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit
title_short Numerical Method on Blood Flow Pressure Drop in Aorta Coronary Sinus Conduit
title_sort numerical method on blood flow pressure drop in aorta coronary sinus conduit
topic Algebra - Graphic methods
Aorta
Sinusitis
url http://psasir.upm.edu.my/id/eprint/9402/
http://psasir.upm.edu.my/id/eprint/9402/1/ID%209402.pdf