Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics

We report the successful use of nanoDots Optically Stimulated Luminescence Dosimeters (OSLDs) for the determination of the entrance surface dose (ESD) in common X-ray diagnostics. ESD evaluation is made on six radiographic projections including abdomen (AP and LAT), chest (AP and PA), skull (AP), an...

Full description

Bibliographic Details
Main Authors: Musa, Yahaya, Hashim, Suhairul, Ghoshal, Sib Krishna, Ahmad, Nurulshuhada Ezzaty, Bradley, D. A., Abdul Karim, Muhammad Khalis, Sabarudin, Akmal
Format: Article
Language:English
Published: Elsevier 2019
Online Access:http://psasir.upm.edu.my/id/eprint/79565/
http://psasir.upm.edu.my/id/eprint/79565/1/Effectiveness%20of%20Al2O3.pdf
_version_ 1848858709498789888
author Musa, Yahaya
Hashim, Suhairul
Ghoshal, Sib Krishna
Ahmad, Nurulshuhada Ezzaty
Bradley, D. A.
Abdul Karim, Muhammad Khalis
Sabarudin, Akmal
author_facet Musa, Yahaya
Hashim, Suhairul
Ghoshal, Sib Krishna
Ahmad, Nurulshuhada Ezzaty
Bradley, D. A.
Abdul Karim, Muhammad Khalis
Sabarudin, Akmal
author_sort Musa, Yahaya
building UPM Institutional Repository
collection Online Access
description We report the successful use of nanoDots Optically Stimulated Luminescence Dosimeters (OSLDs) for the determination of the entrance surface dose (ESD) in common X-ray diagnostics. ESD evaluation is made on six radiographic projections including abdomen (AP and LAT), chest (AP and PA), skull (AP), and thoracic spine (AP). Indirect and direct ESD measurements have been performed with and without whole-body anthropomorphic phantom using GE radiography system installed at the Diagnostic Imaging Laboratory, Universiti Kebangsaan Malaysia (UKM), wherein the International Atomic Energy Agency protocol (IAEA TRS-457) has been followed. The CALDose_X 5.0 software is used to calculate ESDs in these projections. The mean ESDs obtained with indirect measurements for AP abdomen, LAT abdomen, AP chest, PA chest, AP thoracic spine, and AP skull projections are 4.8, 8.3, 0.7, 0.2, 7.2, and 3.7 mGy respectively, and with direct measurements, the respective mean ESDs are 4.3, 8.3, 0.7, 0.3, 6.8, and 2.8 mGy. The values obtained by CALDose_X calculations are 8.3, NA, 1.1, 0.7, 8.8, and 4.5 mGy, respectively. Significant variations ranging from 18% to 59% are observed between measured and calculated ESDs, which are attributed to the difference in phantom size and field size. The overall uncertainty for direct measurements has a maximum of ±0.86 mGy compared to ±0.47 mGy for indirect measurement. Measured ESDs are within the IAEA and European Commission (EC) DRLs range, while calculated ESDs are slightly above the measured values. It is concluded that the use of nanoDots OSLD is beneficial for ESD measurement in routine X-ray examinations.
first_indexed 2025-11-15T12:17:45Z
format Article
id upm-79565
institution Universiti Putra Malaysia
institution_category Local University
language English
last_indexed 2025-11-15T12:17:45Z
publishDate 2019
publisher Elsevier
recordtype eprints
repository_type Digital Repository
spelling upm-795652021-06-29T11:28:01Z http://psasir.upm.edu.my/id/eprint/79565/ Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics Musa, Yahaya Hashim, Suhairul Ghoshal, Sib Krishna Ahmad, Nurulshuhada Ezzaty Bradley, D. A. Abdul Karim, Muhammad Khalis Sabarudin, Akmal We report the successful use of nanoDots Optically Stimulated Luminescence Dosimeters (OSLDs) for the determination of the entrance surface dose (ESD) in common X-ray diagnostics. ESD evaluation is made on six radiographic projections including abdomen (AP and LAT), chest (AP and PA), skull (AP), and thoracic spine (AP). Indirect and direct ESD measurements have been performed with and without whole-body anthropomorphic phantom using GE radiography system installed at the Diagnostic Imaging Laboratory, Universiti Kebangsaan Malaysia (UKM), wherein the International Atomic Energy Agency protocol (IAEA TRS-457) has been followed. The CALDose_X 5.0 software is used to calculate ESDs in these projections. The mean ESDs obtained with indirect measurements for AP abdomen, LAT abdomen, AP chest, PA chest, AP thoracic spine, and AP skull projections are 4.8, 8.3, 0.7, 0.2, 7.2, and 3.7 mGy respectively, and with direct measurements, the respective mean ESDs are 4.3, 8.3, 0.7, 0.3, 6.8, and 2.8 mGy. The values obtained by CALDose_X calculations are 8.3, NA, 1.1, 0.7, 8.8, and 4.5 mGy, respectively. Significant variations ranging from 18% to 59% are observed between measured and calculated ESDs, which are attributed to the difference in phantom size and field size. The overall uncertainty for direct measurements has a maximum of ±0.86 mGy compared to ±0.47 mGy for indirect measurement. Measured ESDs are within the IAEA and European Commission (EC) DRLs range, while calculated ESDs are slightly above the measured values. It is concluded that the use of nanoDots OSLD is beneficial for ESD measurement in routine X-ray examinations. Elsevier 2019-12 Article PeerReviewed text en http://psasir.upm.edu.my/id/eprint/79565/1/Effectiveness%20of%20Al2O3.pdf Musa, Yahaya and Hashim, Suhairul and Ghoshal, Sib Krishna and Ahmad, Nurulshuhada Ezzaty and Bradley, D. A. and Abdul Karim, Muhammad Khalis and Sabarudin, Akmal (2019) Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics. Radiation Physics and Chemistry, 165. ISSN 0969-806X https://www.sciencedirect.com/science/article/abs/pii/S0969806X18301051 10.1016/j.radphyschem.2019.108418
spellingShingle Musa, Yahaya
Hashim, Suhairul
Ghoshal, Sib Krishna
Ahmad, Nurulshuhada Ezzaty
Bradley, D. A.
Abdul Karim, Muhammad Khalis
Sabarudin, Akmal
Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics
title Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics
title_full Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics
title_fullStr Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics
title_full_unstemmed Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics
title_short Effectiveness of Al2O3: C OSL dosimeter towards entrance surface dose measurement in common X-ray diagnostics
title_sort effectiveness of al2o3: c osl dosimeter towards entrance surface dose measurement in common x-ray diagnostics
url http://psasir.upm.edu.my/id/eprint/79565/
http://psasir.upm.edu.my/id/eprint/79565/
http://psasir.upm.edu.my/id/eprint/79565/
http://psasir.upm.edu.my/id/eprint/79565/1/Effectiveness%20of%20Al2O3.pdf