Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter

We examine the influence of elevated dopant concentration on the thermoluminescence characteristics of novel Ge-doped silica fibres. Basic dosimetric characteristics of the TL media were obtained, including linearity, reproducibility, energy dependence, fading, minimum detectable dose and glow curve...

Full description

Bibliographic Details
Main Authors: Mohd Noor, Noramaliza, Ahmad Fadzil, Muhammad Safwan, Ung, Ngie Min, Maah, Mohd Jamil, Mahdiraji, Ghafour Amouzad, Abdul Rashid, Hairul Azhar, Bradley, David A.
Format: Article
Language:English
Published: Elsevier 2016
Online Access:http://psasir.upm.edu.my/id/eprint/53719/
http://psasir.upm.edu.my/id/eprint/53719/1/Radiotherapy%20dosimetry%20and%20the%20thermoluminescence%20characteristics%20of%20Ge-doped%20fibres%20of%20differing%20germanium%20dopant%20concentration%20and%20outer%20diameter.pdf
_version_ 1848852358349455360
author Mohd Noor, Noramaliza
Ahmad Fadzil, Muhammad Safwan
Ung, Ngie Min
Maah, Mohd Jamil
Mahdiraji, Ghafour Amouzad
Abdul Rashid, Hairul Azhar
Bradley, David A.
author_facet Mohd Noor, Noramaliza
Ahmad Fadzil, Muhammad Safwan
Ung, Ngie Min
Maah, Mohd Jamil
Mahdiraji, Ghafour Amouzad
Abdul Rashid, Hairul Azhar
Bradley, David A.
author_sort Mohd Noor, Noramaliza
building UPM Institutional Repository
collection Online Access
description We examine the influence of elevated dopant concentration on the thermoluminescence characteristics of novel Ge-doped silica fibres. Basic dosimetric characteristics of the TL media were obtained, including linearity, reproducibility, energy dependence, fading, minimum detectable dose and glow curve analysis, use being made of a 60Co gamma irradiation facility (mean energy 1.25 MeV) and an electron linear accelerator producing photons at an accelerating potential of 6 and 10 MV. The 6 mol% Ge-doped fibres were found to provide TL response superior to that of 8- and 10 mol% Ge-doped fibres, both for fibres with outer diameter of 241 µm and 604 µm. Concerning reproducibility, obtained under three different test conditions, at <10% the 6 mol% Ge dopant concentration was observed to provide the superior coefficient of variation (CV). In regard to energy dependence, the 10 mol% Ge doped cylindrical fibres produced the largest gradient values at 0.364 and 0.327 for the 241 µm and 604 µm diameter cylindrical fibres respectively and thus the greatest energy dependency. Measured 33 days post irradiation; the 6 mol% Ge doped cylindrical fibres showed the least TL signal loss, at 21% for the 241 µm cylindrical fibre and <40% for the 604 µm cylindrical fibres. The results also revealed that the 6 mol% optical fibres provided the lowest minimum detectable dose, at 0.027 Gy for 6 MV photon beams. Evaluations of these characteristics are supporting development of novel Ge-doped optical fibres for dosimetry in radiotherapy.
first_indexed 2025-11-15T10:36:48Z
format Article
id upm-53719
institution Universiti Putra Malaysia
institution_category Local University
language English
last_indexed 2025-11-15T10:36:48Z
publishDate 2016
publisher Elsevier
recordtype eprints
repository_type Digital Repository
spelling upm-537192018-01-15T06:15:45Z http://psasir.upm.edu.my/id/eprint/53719/ Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter Mohd Noor, Noramaliza Ahmad Fadzil, Muhammad Safwan Ung, Ngie Min Maah, Mohd Jamil Mahdiraji, Ghafour Amouzad Abdul Rashid, Hairul Azhar Bradley, David A. We examine the influence of elevated dopant concentration on the thermoluminescence characteristics of novel Ge-doped silica fibres. Basic dosimetric characteristics of the TL media were obtained, including linearity, reproducibility, energy dependence, fading, minimum detectable dose and glow curve analysis, use being made of a 60Co gamma irradiation facility (mean energy 1.25 MeV) and an electron linear accelerator producing photons at an accelerating potential of 6 and 10 MV. The 6 mol% Ge-doped fibres were found to provide TL response superior to that of 8- and 10 mol% Ge-doped fibres, both for fibres with outer diameter of 241 µm and 604 µm. Concerning reproducibility, obtained under three different test conditions, at <10% the 6 mol% Ge dopant concentration was observed to provide the superior coefficient of variation (CV). In regard to energy dependence, the 10 mol% Ge doped cylindrical fibres produced the largest gradient values at 0.364 and 0.327 for the 241 µm and 604 µm diameter cylindrical fibres respectively and thus the greatest energy dependency. Measured 33 days post irradiation; the 6 mol% Ge doped cylindrical fibres showed the least TL signal loss, at 21% for the 241 µm cylindrical fibre and <40% for the 604 µm cylindrical fibres. The results also revealed that the 6 mol% optical fibres provided the lowest minimum detectable dose, at 0.027 Gy for 6 MV photon beams. Evaluations of these characteristics are supporting development of novel Ge-doped optical fibres for dosimetry in radiotherapy. Elsevier 2016-09 Article PeerReviewed application/pdf en http://psasir.upm.edu.my/id/eprint/53719/1/Radiotherapy%20dosimetry%20and%20the%20thermoluminescence%20characteristics%20of%20Ge-doped%20fibres%20of%20differing%20germanium%20dopant%20concentration%20and%20outer%20diameter.pdf Mohd Noor, Noramaliza and Ahmad Fadzil, Muhammad Safwan and Ung, Ngie Min and Maah, Mohd Jamil and Mahdiraji, Ghafour Amouzad and Abdul Rashid, Hairul Azhar and Bradley, David A. (2016) Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter. Radiation Physics and Chemistry, 126. pp. 56-61. ISSN 0969-806X; ESSN: 1879-0895 https://www.sciencedirect.com/science/article/pii/S0969806X16301347 10.1016/j.radphyschem.2016.05.001
spellingShingle Mohd Noor, Noramaliza
Ahmad Fadzil, Muhammad Safwan
Ung, Ngie Min
Maah, Mohd Jamil
Mahdiraji, Ghafour Amouzad
Abdul Rashid, Hairul Azhar
Bradley, David A.
Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter
title Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter
title_full Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter
title_fullStr Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter
title_full_unstemmed Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter
title_short Radiotherapy dosimetry and the thermoluminescence characteristics of Ge-doped fibres of differing germanium dopant concentration and outer diameter
title_sort radiotherapy dosimetry and the thermoluminescence characteristics of ge-doped fibres of differing germanium dopant concentration and outer diameter
url http://psasir.upm.edu.my/id/eprint/53719/
http://psasir.upm.edu.my/id/eprint/53719/
http://psasir.upm.edu.my/id/eprint/53719/
http://psasir.upm.edu.my/id/eprint/53719/1/Radiotherapy%20dosimetry%20and%20the%20thermoluminescence%20characteristics%20of%20Ge-doped%20fibres%20of%20differing%20germanium%20dopant%20concentration%20and%20outer%20diameter.pdf