Low power upconversion mixer for medical remote sensing
This work presents the design of a low power upconversion mixer adapted in medical remote sensing such as wireless endoscopy application. The proposed upconversion mixer operates in ISM band of 433 MHz. With the carrier power of −5 dBm, the proposed mixer has an output inferred 1 dB compression poin...
| Main Authors: | , , , |
|---|---|
| Format: | Article |
| Published: |
Hindawi Publishing Corporation
2014
|
| Online Access: | http://psasir.upm.edu.my/id/eprint/34402/ |
| _version_ | 1848847762670485504 |
|---|---|
| author | Lioe, De Xing Shafie, Suhaidi Ramiah, Harikrishnan Tan, Gim Heng |
| author_facet | Lioe, De Xing Shafie, Suhaidi Ramiah, Harikrishnan Tan, Gim Heng |
| author_sort | Lioe, De Xing |
| building | UPM Institutional Repository |
| collection | Online Access |
| description | This work presents the design of a low power upconversion mixer adapted in medical remote sensing such as wireless endoscopy application. The proposed upconversion mixer operates in ISM band of 433 MHz. With the carrier power of −5 dBm, the proposed mixer has an output inferred 1 dB compression point of −0.5 dBm with a corresponding output third-order intercept point (OIP3) of 7.1 dBm. The design of the upconversion mixer is realized on CMOS 0.13 μm platform, with a current consumption of 594 μA at supply voltage headroom of 1.2 V. |
| first_indexed | 2025-11-15T09:23:46Z |
| format | Article |
| id | upm-34402 |
| institution | Universiti Putra Malaysia |
| institution_category | Local University |
| last_indexed | 2025-11-15T09:23:46Z |
| publishDate | 2014 |
| publisher | Hindawi Publishing Corporation |
| recordtype | eprints |
| repository_type | Digital Repository |
| spelling | upm-344022015-12-15T00:57:37Z http://psasir.upm.edu.my/id/eprint/34402/ Low power upconversion mixer for medical remote sensing Lioe, De Xing Shafie, Suhaidi Ramiah, Harikrishnan Tan, Gim Heng This work presents the design of a low power upconversion mixer adapted in medical remote sensing such as wireless endoscopy application. The proposed upconversion mixer operates in ISM band of 433 MHz. With the carrier power of −5 dBm, the proposed mixer has an output inferred 1 dB compression point of −0.5 dBm with a corresponding output third-order intercept point (OIP3) of 7.1 dBm. The design of the upconversion mixer is realized on CMOS 0.13 μm platform, with a current consumption of 594 μA at supply voltage headroom of 1.2 V. Hindawi Publishing Corporation 2014 Article PeerReviewed Lioe, De Xing and Shafie, Suhaidi and Ramiah, Harikrishnan and Tan, Gim Heng (2014) Low power upconversion mixer for medical remote sensing. The Scientific World Journal, 2014. art. no. 923893. pp. 1-5. ISSN 2356-6140; ESSN: 1537-744X http://www.hindawi.com/journals/tswj/2014/923893/abs/ 10.1155/2014/923893 |
| spellingShingle | Lioe, De Xing Shafie, Suhaidi Ramiah, Harikrishnan Tan, Gim Heng Low power upconversion mixer for medical remote sensing |
| title | Low power upconversion mixer for medical remote sensing |
| title_full | Low power upconversion mixer for medical remote sensing |
| title_fullStr | Low power upconversion mixer for medical remote sensing |
| title_full_unstemmed | Low power upconversion mixer for medical remote sensing |
| title_short | Low power upconversion mixer for medical remote sensing |
| title_sort | low power upconversion mixer for medical remote sensing |
| url | http://psasir.upm.edu.my/id/eprint/34402/ http://psasir.upm.edu.my/id/eprint/34402/ http://psasir.upm.edu.my/id/eprint/34402/ |