Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy

Fourier transform infrared (FTIR) spectroscopy with transmission cell is described to predict anisidine value of palm olein. The calibration set was prepared by mixing the thermally oxidized palm olein and the unoxidized palm olein with certain ratios (w/w) covering a wide range of anisidine values....

Full description

Bibliographic Details
Main Authors: Che Man, Y. B., Setiowaty, G.
Format: Article
Language:English
Published: Wiley 1999
Online Access:http://psasir.upm.edu.my/id/eprint/112639/
http://psasir.upm.edu.my/id/eprint/112639/3/112639.pdf
_version_ 1848865997328482304
author Che Man, Y. B.
Setiowaty, G.
author_facet Che Man, Y. B.
Setiowaty, G.
author_sort Che Man, Y. B.
building UPM Institutional Repository
collection Online Access
description Fourier transform infrared (FTIR) spectroscopy with transmission cell is described to predict anisidine value of palm olein. The calibration set was prepared by mixing the thermally oxidized palm olein and the unoxidized palm olein with certain ratios (w/w) covering a wide range of anisidine values. A partial least square (PLS) regression technique was employed to construct a calibration model. This model was further accomplished by a validation step. The standard error of prediction found was 0.51. The precision of this method was shown to be comparable to the accuracy of the American Oil Chemists’ Society method used for measurement of anisidine value, with coefficient of determination (R2) of 0.99. The study showed that mid‐band FTIR spectroscopy combined with a PLS calibration technique is a versatile, efficient, and accurate technique for the estimation of anisidine value of palm olein within about 2 min with less than 2 mL of sample.
first_indexed 2025-11-15T14:13:36Z
format Article
id upm-112639
institution Universiti Putra Malaysia
institution_category Local University
language English
last_indexed 2025-11-15T14:13:36Z
publishDate 1999
publisher Wiley
recordtype eprints
repository_type Digital Repository
spelling upm-1126392025-03-06T03:14:32Z http://psasir.upm.edu.my/id/eprint/112639/ Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy Che Man, Y. B. Setiowaty, G. Fourier transform infrared (FTIR) spectroscopy with transmission cell is described to predict anisidine value of palm olein. The calibration set was prepared by mixing the thermally oxidized palm olein and the unoxidized palm olein with certain ratios (w/w) covering a wide range of anisidine values. A partial least square (PLS) regression technique was employed to construct a calibration model. This model was further accomplished by a validation step. The standard error of prediction found was 0.51. The precision of this method was shown to be comparable to the accuracy of the American Oil Chemists’ Society method used for measurement of anisidine value, with coefficient of determination (R2) of 0.99. The study showed that mid‐band FTIR spectroscopy combined with a PLS calibration technique is a versatile, efficient, and accurate technique for the estimation of anisidine value of palm olein within about 2 min with less than 2 mL of sample. Wiley 1999 Article PeerReviewed text en http://psasir.upm.edu.my/id/eprint/112639/3/112639.pdf Che Man, Y. B. and Setiowaty, G. (1999) Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy. Journal of the American Oil Chemists' Society, 76 (2). pp. 243-247. ISSN 0003-021X; eISSN: 1558-9331 https://aocs.onlinelibrary.wiley.com/doi/10.1007/s11746-999-0225-x 10.1007/s11746-999-0225-x
spellingShingle Che Man, Y. B.
Setiowaty, G.
Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy
title Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy
title_full Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy
title_fullStr Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy
title_full_unstemmed Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy
title_short Determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy
title_sort determination of anisidine value in thermally oxidized palm olein by fourier transform infrared spectroscopy
url http://psasir.upm.edu.my/id/eprint/112639/
http://psasir.upm.edu.my/id/eprint/112639/
http://psasir.upm.edu.my/id/eprint/112639/
http://psasir.upm.edu.my/id/eprint/112639/3/112639.pdf