Determination of iodine value of palm oil by fourier transform infrared spectroscopy

A rapid method for the quantitative determination of iodine value (IV) of palm oil products by FTIR transmission spectroscopy is described. A calibration standard was developed by blending palm stearin and superolein in specific ratios that covered a range of 27.9 to 65.3 IV units. The spectra of th...

Full description

Bibliographic Details
Main Authors: Che Man, Y. B., Setiowaty, G., van de Voort, F. R.
Format: Article
Language:English
Published: Wiley 1999
Online Access:http://psasir.upm.edu.my/id/eprint/111937/
http://psasir.upm.edu.my/id/eprint/111937/3/111937.pdf
Description
Summary:A rapid method for the quantitative determination of iodine value (IV) of palm oil products by FTIR transmission spectroscopy is described. A calibration standard was developed by blending palm stearin and superolein in specific ratios that covered a range of 27.9 to 65.3 IV units. The spectra of these standards was measured in the range between 3050 and 2984 cm−1, corresponding to the absorption band of=C-H cis stretching vibration. A partial least squares calibration model for the prediction of IV was developed to quantify the IV of palm oil products. A validation approach was used to optimize the calibration with a correlation coefficient of R2=0.9995 and a standard error of prediction of 0.151. This study concludes that the FTIR transmission approach can be used to determine the IV of palm oil products with a total analysis time per sample of less than 2 min for liquid samples.