Activity determination of 137Cs in NPL’s ampoule using a point source
The activity of two liquid mixed radionuclide sources in NPL'S standard glass ampoules were determined by gamma-ray spectrometry measurement using a standard point source placed at 38.5 cm axially above the detector. A radio nuclide 137Cs was chosen, Results obtained were within 3% agreement wi...
| Main Author: | |
|---|---|
| Format: | Article |
| Published: |
Universiti Kebangsaan Malaysia
1997
|
| Online Access: | http://journalarticle.ukm.my/3756/ |
| _version_ | 1848810292966850560 |
|---|---|
| author | Supian Bin Samat, |
| author_facet | Supian Bin Samat, |
| author_sort | Supian Bin Samat, |
| building | UKM Institutional Repository |
| collection | Online Access |
| description | The activity of two liquid mixed radionuclide sources in NPL'S standard glass ampoules were determined by gamma-ray spectrometry measurement using a standard point source placed at 38.5 cm axially above the detector. A radio nuclide 137Cs was chosen, Results obtained were within 3% agreement with the calculated value. The same degree of deviation was obtained when the NPL standard source of equal geometry, chemical composition and density was used. |
| first_indexed | 2025-11-14T23:28:12Z |
| format | Article |
| id | oai:generic.eprints.org:3756 |
| institution | Universiti Kebangasaan Malaysia |
| institution_category | Local University |
| last_indexed | 2025-11-14T23:28:12Z |
| publishDate | 1997 |
| publisher | Universiti Kebangsaan Malaysia |
| recordtype | eprints |
| repository_type | Digital Repository |
| spelling | oai:generic.eprints.org:37562012-03-19T03:10:59Z http://journalarticle.ukm.my/3756/ Activity determination of 137Cs in NPL’s ampoule using a point source Supian Bin Samat, The activity of two liquid mixed radionuclide sources in NPL'S standard glass ampoules were determined by gamma-ray spectrometry measurement using a standard point source placed at 38.5 cm axially above the detector. A radio nuclide 137Cs was chosen, Results obtained were within 3% agreement with the calculated value. The same degree of deviation was obtained when the NPL standard source of equal geometry, chemical composition and density was used. Universiti Kebangsaan Malaysia 1997-09 Article PeerReviewed Supian Bin Samat, (1997) Activity determination of 137Cs in NPL’s ampoule using a point source. Sains Malaysiana, 26 (3&4). pp. 81-86. ISSN 0126-6039 http://www.ukm.my/jsm/english_journals/vol26num3and4_1997/vol26num3and4_97page81-86.html |
| spellingShingle | Supian Bin Samat, Activity determination of 137Cs in NPL’s ampoule using a point source |
| title | Activity determination of 137Cs in NPL’s ampoule using a point source |
| title_full | Activity determination of 137Cs in NPL’s ampoule using a point source |
| title_fullStr | Activity determination of 137Cs in NPL’s ampoule using a point source |
| title_full_unstemmed | Activity determination of 137Cs in NPL’s ampoule using a point source |
| title_short | Activity determination of 137Cs in NPL’s ampoule using a point source |
| title_sort | activity determination of 137cs in npl’s ampoule using a point source |
| url | http://journalarticle.ukm.my/3756/ http://journalarticle.ukm.my/3756/ |