Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode

An ultraviolet A (UVA) radiation intensity sensor with responsivity in the wavelength range of 320-360 nm was developed based on a gallium nitride (GaN) photodiode. In this sensor system, a GaN photodiode in reverse-biased mode converts the radiation intensity into current, which was then converted...

Full description

Bibliographic Details
Main Authors: Theyirakumar .J, Gopir .G, Yatim .B, Sanusi .H, Megat Mahmud .P.S, Hoe .P.C
Format: Article
Language:English
Published: Universiti Kebangsaan Malaysia 2011
Online Access:http://journalarticle.ukm.my/2441/
http://journalarticle.ukm.my/2441/1/06_Theyirakumar.pdf
_version_ 1848809867227168768
author Theyirakumar .J,
Gopir .G,
Yatim .B,
Sanusi .H,
Megat Mahmud .P.S,
Hoe .P.C,
author_facet Theyirakumar .J,
Gopir .G,
Yatim .B,
Sanusi .H,
Megat Mahmud .P.S,
Hoe .P.C,
author_sort Theyirakumar .J,
building UKM Institutional Repository
collection Online Access
description An ultraviolet A (UVA) radiation intensity sensor with responsivity in the wavelength range of 320-360 nm was developed based on a gallium nitride (GaN) photodiode. In this sensor system, a GaN photodiode in reverse-biased mode converts the radiation intensity into current, which was then converted and amplified into an output voltage by a transimpedance amplifier (TIA), or current-voltage converter, consisting of an operational amplifier and a feedback resistor. For a narrowband UV source, the radiation intensity could be calculated from the values of the output voltage, feedback resistor, photodiode responsivity and photodiode effective area. The sensor was tested by performing measurements over different values of UV source wavelength, source distance, ambient temperature and sampling time. For calibration with a broadband UV source, the GaN-UVA sensor was used simultaneously with a standard Si-UVA sensor to measure solar radiation. The observed linear relationship between the sensors’ outputs enables us to convert the output voltage of the GaN-UVA sensor to UVA intensity. Thus, we have successfully developed, tested and calibrated an ultraviolet A radiation sensor based on the GaN photodiode.
first_indexed 2025-11-14T23:21:26Z
format Article
id oai:generic.eprints.org:2441
institution Universiti Kebangasaan Malaysia
institution_category Local University
language English
last_indexed 2025-11-14T23:21:26Z
publishDate 2011
publisher Universiti Kebangsaan Malaysia
recordtype eprints
repository_type Digital Repository
spelling oai:generic.eprints.org:24412016-12-14T06:31:37Z http://journalarticle.ukm.my/2441/ Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode Theyirakumar .J, Gopir .G, Yatim .B, Sanusi .H, Megat Mahmud .P.S, Hoe .P.C, An ultraviolet A (UVA) radiation intensity sensor with responsivity in the wavelength range of 320-360 nm was developed based on a gallium nitride (GaN) photodiode. In this sensor system, a GaN photodiode in reverse-biased mode converts the radiation intensity into current, which was then converted and amplified into an output voltage by a transimpedance amplifier (TIA), or current-voltage converter, consisting of an operational amplifier and a feedback resistor. For a narrowband UV source, the radiation intensity could be calculated from the values of the output voltage, feedback resistor, photodiode responsivity and photodiode effective area. The sensor was tested by performing measurements over different values of UV source wavelength, source distance, ambient temperature and sampling time. For calibration with a broadband UV source, the GaN-UVA sensor was used simultaneously with a standard Si-UVA sensor to measure solar radiation. The observed linear relationship between the sensors’ outputs enables us to convert the output voltage of the GaN-UVA sensor to UVA intensity. Thus, we have successfully developed, tested and calibrated an ultraviolet A radiation sensor based on the GaN photodiode. Universiti Kebangsaan Malaysia 2011-01 Article PeerReviewed application/pdf en http://journalarticle.ukm.my/2441/1/06_Theyirakumar.pdf Theyirakumar .J, and Gopir .G, and Yatim .B, and Sanusi .H, and Megat Mahmud .P.S, and Hoe .P.C, (2011) Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode. Sains Malaysiana, 40 (1). pp. 21-25. ISSN 0126-6039 http://www.ukm.my/jsm/
spellingShingle Theyirakumar .J,
Gopir .G,
Yatim .B,
Sanusi .H,
Megat Mahmud .P.S,
Hoe .P.C,
Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode
title Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode
title_full Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode
title_fullStr Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode
title_full_unstemmed Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode
title_short Testing and calibration of an ultraviolet-a radiation sensor based on GaN photodiode
title_sort testing and calibration of an ultraviolet-a radiation sensor based on gan photodiode
url http://journalarticle.ukm.my/2441/
http://journalarticle.ukm.my/2441/
http://journalarticle.ukm.my/2441/1/06_Theyirakumar.pdf