Ultra-high frequency magnetic resonance imaging

This thesis addresses the problem of radiofrequency probe design for Ultra High Frequency Magnetic Resonance Imaging (7T). The signal-to-noise ratio available in Magnetic Resonance Imaging (MRI) is determined by the static magnetic field strength, causing a continued drive toward higher fields to e...

Full description

Bibliographic Details
Main Author: Magill, Arthur W.
Format: Thesis (University of Nottingham only)
Language:English
Published: 2007
Subjects:
Online Access:https://eprints.nottingham.ac.uk/10740/
_version_ 1848791121990254592
author Magill, Arthur W.
author_facet Magill, Arthur W.
author_sort Magill, Arthur W.
building Nottingham Research Data Repository
collection Online Access
description This thesis addresses the problem of radiofrequency probe design for Ultra High Frequency Magnetic Resonance Imaging (7T). The signal-to-noise ratio available in Magnetic Resonance Imaging (MRI) is determined by the static magnetic field strength, causing a continued drive toward higher fields to enable faster image acquisition at finer spatial resolution. The resonant frequency increases linearly with static field strength. At 7T the proton resonant frequency is 300MHz, with a wavelength of approximately 13cm in tissue. As this is smaller than the dimensions of the human head, the phase of the radiofrequency (RF) signal varies considerably across the sample, producing field cancellation due to interference. A full wave electromagnetic simulator, using the Transmission Line Matrix (TLM) method, was developed to investigate RF probes at high frequency. A Birdcage probe operating at 64, 128 and 300MHz (corresponding to 1.5, 3 and 7T) was simulated, loaded with an anatomically detailed human head model. A half-wave microstrip was investigated for use as a high frequency probe element. Magnetic and electric fields produced by a single microstrip were simulated, and the strip dimensions varied to investigate the effect on field penetration into the head and Specific Absorption Rate (SAR). A transmit-receive array probe using four microstrip elements was then developed. Bloch simulations were run, using TLM generated magnetic fields, to investigate imaging at short wavelength. Parallel receive probes are demonstrated to offer considerable advantage over volume probes, as signals from receive elements can be combined without interference. There is no transmit equivalent to parallel reception; simultaneous excitation of independent probe elements causes interference in exactly the same manner as a volume probe. A new imaging sequence was developed using a Burst-like encoding to allow sequential excitation of probe elements, without interference, which can be recalled in a single readout. An improvement in image homogeneity was demonstrated, and SENSE acceleration of the new imaging sequence is shown. The sequence was implemented at 3T using a purpose built four element microstrip probe. An RF multiplexer was also built to enable transmit element switching during the imaging sequence. It was demonstrated that images due to different RF excitations, acquired in a single EPI readout, can be separated.
first_indexed 2025-11-14T18:23:29Z
format Thesis (University of Nottingham only)
id nottingham-10740
institution University of Nottingham Malaysia Campus
institution_category Local University
language English
last_indexed 2025-11-14T18:23:29Z
publishDate 2007
recordtype eprints
repository_type Digital Repository
spelling nottingham-107402025-02-28T11:09:23Z https://eprints.nottingham.ac.uk/10740/ Ultra-high frequency magnetic resonance imaging Magill, Arthur W. This thesis addresses the problem of radiofrequency probe design for Ultra High Frequency Magnetic Resonance Imaging (7T). The signal-to-noise ratio available in Magnetic Resonance Imaging (MRI) is determined by the static magnetic field strength, causing a continued drive toward higher fields to enable faster image acquisition at finer spatial resolution. The resonant frequency increases linearly with static field strength. At 7T the proton resonant frequency is 300MHz, with a wavelength of approximately 13cm in tissue. As this is smaller than the dimensions of the human head, the phase of the radiofrequency (RF) signal varies considerably across the sample, producing field cancellation due to interference. A full wave electromagnetic simulator, using the Transmission Line Matrix (TLM) method, was developed to investigate RF probes at high frequency. A Birdcage probe operating at 64, 128 and 300MHz (corresponding to 1.5, 3 and 7T) was simulated, loaded with an anatomically detailed human head model. A half-wave microstrip was investigated for use as a high frequency probe element. Magnetic and electric fields produced by a single microstrip were simulated, and the strip dimensions varied to investigate the effect on field penetration into the head and Specific Absorption Rate (SAR). A transmit-receive array probe using four microstrip elements was then developed. Bloch simulations were run, using TLM generated magnetic fields, to investigate imaging at short wavelength. Parallel receive probes are demonstrated to offer considerable advantage over volume probes, as signals from receive elements can be combined without interference. There is no transmit equivalent to parallel reception; simultaneous excitation of independent probe elements causes interference in exactly the same manner as a volume probe. A new imaging sequence was developed using a Burst-like encoding to allow sequential excitation of probe elements, without interference, which can be recalled in a single readout. An improvement in image homogeneity was demonstrated, and SENSE acceleration of the new imaging sequence is shown. The sequence was implemented at 3T using a purpose built four element microstrip probe. An RF multiplexer was also built to enable transmit element switching during the imaging sequence. It was demonstrated that images due to different RF excitations, acquired in a single EPI readout, can be separated. 2007-07-12 Thesis (University of Nottingham only) NonPeerReviewed application/pdf en arr https://eprints.nottingham.ac.uk/10740/1/magill_thesis.pdf Magill, Arthur W. (2007) Ultra-high frequency magnetic resonance imaging. PhD thesis, University of Nottingham. Magnetic Resonance Imaging MRI Ultra High Field 7 Tesla Radiofrequency coil design Electromagnetic simulation Transmission Line Matrix TLM Hardware
spellingShingle Magnetic Resonance Imaging MRI Ultra High Field 7 Tesla Radiofrequency coil design Electromagnetic simulation Transmission Line Matrix TLM Hardware
Magill, Arthur W.
Ultra-high frequency magnetic resonance imaging
title Ultra-high frequency magnetic resonance imaging
title_full Ultra-high frequency magnetic resonance imaging
title_fullStr Ultra-high frequency magnetic resonance imaging
title_full_unstemmed Ultra-high frequency magnetic resonance imaging
title_short Ultra-high frequency magnetic resonance imaging
title_sort ultra-high frequency magnetic resonance imaging
topic Magnetic Resonance Imaging MRI Ultra High Field 7 Tesla Radiofrequency coil design Electromagnetic simulation Transmission Line Matrix TLM Hardware
url https://eprints.nottingham.ac.uk/10740/