Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography
Introduction: Various medium and high tube potentials were utilized to conduct chest x-rays. There are advantages and disadvantages with regards to image quality and radiation dose when using medium and high kilovoltage (kVp) technique. However, radiographers have misconstrued understanding pertain...
| Main Authors: | Moey, Soo Foon, Muhd Jaafar, Nur Farah Hani, Saidin, Nursyahirah |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Kulliyyah of Allied Health Sciences, IIUM
2019
|
| Subjects: | |
| Online Access: | http://irep.iium.edu.my/82902/ http://irep.iium.edu.my/82902/1/274-Article%20Text-1915-1-10-20200304%20%282%29%20Farah%20Hanim.pdf |
Similar Items
Effects of kilovoltage on image quality and entrance surface dose in lumbar spine digital radiography
by: Ramli, Hanis Aisyah, et al.
Published: (2018)
by: Ramli, Hanis Aisyah, et al.
Published: (2018)
Effects of kilovoltage on image quality and entrance surface dose in lumbar spine digital radiography
by: Ramli, Hanis Aisyah, et al.
Published: (2019)
by: Ramli, Hanis Aisyah, et al.
Published: (2019)
The practice of chest radiography using different digital imaging systems: Dose and image quality
by: Moey, Soo Foon, et al.
Published: (2018)
by: Moey, Soo Foon, et al.
Published: (2018)
A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
by: Moey, Soo Foon, et al.
Published: (2018)
by: Moey, Soo Foon, et al.
Published: (2018)
Image optimisation and dose reduction in computed radiography: pa chest and contrast detail phantom
by: Ahmad Radzi, Halimatussa’diah, et al.
Published: (2015)
by: Ahmad Radzi, Halimatussa’diah, et al.
Published: (2015)
Optimisation of image quality and radiation dose using air-gap technique in pelvic radiography
by: Moey, Soo Foon, et al.
Published: (2016)
by: Moey, Soo Foon, et al.
Published: (2016)
Computed and conventional chest radiography: A comparison of image quality and radiation dose
by: Ramli, Khadijah, et al.
Published: (2005)
by: Ramli, Khadijah, et al.
Published: (2005)
Investigation of imaging parameters and radiation dose in chest radiographic imaging: A comparison of computed radiography and direct radiography
by: Sun, Zhonghua, et al.
Published: (2008)
by: Sun, Zhonghua, et al.
Published: (2008)
Optimization of dose and image quality in full field and computed radiography systems for common digital radiographic examinations
by: Moey, Soo Foon, et al.
Published: (2018)
by: Moey, Soo Foon, et al.
Published: (2018)
Is the 28-day rule safe for use in abdominal radiography?
by: Moey, Soo Foon, et al.
Published: (2020)
by: Moey, Soo Foon, et al.
Published: (2020)
Computed Radiography of Chest Radiographic Imaging: A Phantom Study of Optimal Imaging Parameters and Radiation Dose
by: Sun, Zhonghua, et al.
Published: (2007)
by: Sun, Zhonghua, et al.
Published: (2007)
Optimization of chest radiographic imaging parameters: a comparison of image quality and entrance skin dose for digital chest radiography systems
by: Sun, Zhonghua, et al.
Published: (2012)
by: Sun, Zhonghua, et al.
Published: (2012)
ABCs of chest radiography
by: Hassan, Radhiana
Published: (2019)
by: Hassan, Radhiana
Published: (2019)
Computed radiography and direct radiography of chest radiographic imaging: a comparision of image quality and entrance skin dose for various systems
by: Sun, Zhonghua, et al.
Published: (2009)
by: Sun, Zhonghua, et al.
Published: (2009)
Evaluation of the influence of exposure index on image quality and radiation dose
by: Moey, Soo Foon, et al.
Published: (2019)
by: Moey, Soo Foon, et al.
Published: (2019)
Radiographic appearance in plain radiography of the chest (CXR)
by: Che Mohamed, Siti Kamariah
Published: (2018)
by: Che Mohamed, Siti Kamariah
Published: (2018)
Scattered radiation dose to caregivers’ in plain radiography: a case study
by: Sayed, Inayatullah Shah, et al.
Published: (2019)
by: Sayed, Inayatullah Shah, et al.
Published: (2019)
Chest computed radiography imaging parameters for nodule detection: a comparison of image quality and entrance skin dose
by: Sun, Zhonghua
Published: (2014)
by: Sun, Zhonghua
Published: (2014)
Optimization of dose and image quality in CR and DR for common digital radiographic examinations
by: Moey, Soo Foon, et al.
Published: (2018)
by: Moey, Soo Foon, et al.
Published: (2018)
A survey on the knowledge of breast cancer and adoption of breast self-examination among female undergraduate students of Kulliyyah of Allied Health Sciences, IIUM Kuantan
by: Saidin, Nursyahirah, et al.
Published: (2018)
by: Saidin, Nursyahirah, et al.
Published: (2018)
Statistical methods for computer aided diagnosis in chest radiography / Hossein Ebrahimian
by: Ebrahimian, Hossein
Published: (2013)
by: Ebrahimian, Hossein
Published: (2013)
Dose evaluation for common digital radiographic examinations in selected hospitals in Pahang Malaysia
by: Moey, Soo Foon, et al.
Published: (2017)
by: Moey, Soo Foon, et al.
Published: (2017)
Optimization of dose and image quality in CR and DR for common digital radiographic examinations
by: Moey, Soo Foon, et al.
Published: (2017)
by: Moey, Soo Foon, et al.
Published: (2017)
Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
by: Abdul Aziz Ismail,, et al.
Published: (2017)
by: Abdul Aziz Ismail,, et al.
Published: (2017)
Modified technique for the visualisation of C6/C7 in the lateral cervical spine radiography
by: Moey, Soo Foon, et al.
Published: (2019)
by: Moey, Soo Foon, et al.
Published: (2019)
Dosimetric Verification Using Optically Stimulated Luminescence (Osld) Nanodots For Postero-Anterior (Pa) Chest Examinations In Perak Health Clinics
by: Saidin, Mohd Tarmizi
Published: (2021)
by: Saidin, Mohd Tarmizi
Published: (2021)
Designing surveys for radiography and radiation therapy.
by: O'Connor, Moira, et al.
Published: (2010)
by: O'Connor, Moira, et al.
Published: (2010)
Effects of educational intervention on repetition rate of routine digital chest radiography among radiographers in tertiary hospitals at Makkah, Saudi Arabia
by: Almalki, Abdullah Ahmed G
Published: (2019)
by: Almalki, Abdullah Ahmed G
Published: (2019)
Investigation of optimal parameters for chest radiograph imaging using film-screen and computed radiography
by: Sun, Zhonghua, et al.
Published: (2007)
by: Sun, Zhonghua, et al.
Published: (2007)
Patient’s effective dose and performance assessment of computed radiography systems
by: Bushra, A., et al.
Published: (2022)
by: Bushra, A., et al.
Published: (2022)
Image quality and dose analysis for a PA chest X-ray: Comparison between AEC mode acquisition and manual mode using the 10 kVp 'rule'
by: Reis, Claudia, et al.
Published: (2014)
by: Reis, Claudia, et al.
Published: (2014)
The effect of varying kilovoltage (kVp) and tube current (mAs) on the image quality and dose of CTA head phantom / Sity Noor Ayseah Dzulkafli
by: Dzulkafli, Sity Noor Ayseah
Published: (2015)
by: Dzulkafli, Sity Noor Ayseah
Published: (2015)
Minimization Of Entrance Surface Dose And Critical Organ Dose For Medical Radiography Using Optimization Procedures
by: Nassab Rabor, Lotf Ali Mehdipour
Published: (2010)
by: Nassab Rabor, Lotf Ali Mehdipour
Published: (2010)
Machine learning-based classification of Covid-19 using chest radiography images / Chan Yu T'ng
by: Chan, Yu T'ng
Published: (2021)
by: Chan, Yu T'ng
Published: (2021)
Evaluation of a Homegrown Computed Radiography Dose Data Mining Program for Online Automatic Surveillance of Dose Creep
by: Ng, Curtise, et al.
Published: (2009)
by: Ng, Curtise, et al.
Published: (2009)
Image quality and entrance surface dose evaluation for
lateral cervical spine: a study using grid and non-grid techniques
by: Moey, Soo Foon, et al.
Published: (2018)
by: Moey, Soo Foon, et al.
Published: (2018)
The determination of absorbed dose
in femur bone marrow during x-ray
radiography through fabrication of
physical and digital femur phantoms
by: Dehyagani, Zaker Salehi
Published: (2013)
by: Dehyagani, Zaker Salehi
Published: (2013)
Educational module intervention for radiographers to reduce repetition rate of routine digital chest radiography in Makkah region of Saudi Arabia tertiary hospitals: protocol of a quasi-experimental study
by: A. Almalki, Abdullah, et al.
Published: (2017)
by: A. Almalki, Abdullah, et al.
Published: (2017)
Repeat rate analysis of lower limb radiograph for digital radiography in general radiography at Pantai Klang Hospital Selangor
Analysis of image quality of pediatric chest radiography in the neonatal intensive care unit (NICU) of Hospital Universiti Sains Malaysia (HUSM)
by: Awal, Mohd Firdaus Mohd
Published: (2023)
by: Awal, Mohd Firdaus Mohd
Published: (2023)
Similar Items
-
Effects of kilovoltage on image quality and entrance surface dose in lumbar spine digital radiography
by: Ramli, Hanis Aisyah, et al.
Published: (2018) -
Effects of kilovoltage on image quality and entrance surface dose in lumbar spine digital radiography
by: Ramli, Hanis Aisyah, et al.
Published: (2019) -
The practice of chest radiography using different digital imaging systems: Dose and image quality
by: Moey, Soo Foon, et al.
Published: (2018) -
A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
by: Moey, Soo Foon, et al.
Published: (2018) -
Image optimisation and dose reduction in computed radiography: pa chest and contrast detail phantom
by: Ahmad Radzi, Halimatussa’diah, et al.
Published: (2015)