Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography

Introduction: Various medium and high tube potentials were utilized to conduct chest x-rays. There are advantages and disadvantages with regards to image quality and radiation dose when using medium and high kilovoltage (kVp) technique. However, radiographers have misconstrued understanding pertain...

Full description

Bibliographic Details
Main Authors: Moey, Soo Foon, Muhd Jaafar, Nur Farah Hani, Saidin, Nursyahirah
Format: Article
Language:English
Published: Kulliyyah of Allied Health Sciences, IIUM 2019
Subjects:
Online Access:http://irep.iium.edu.my/82902/
http://irep.iium.edu.my/82902/1/274-Article%20Text-1915-1-10-20200304%20%282%29%20Farah%20Hanim.pdf
_version_ 1848789388212830208
author Moey, Soo Foon
Muhd Jaafar, Nur Farah Hani
Saidin, Nursyahirah
author_facet Moey, Soo Foon
Muhd Jaafar, Nur Farah Hani
Saidin, Nursyahirah
author_sort Moey, Soo Foon
building IIUM Repository
collection Online Access
description Introduction: Various medium and high tube potentials were utilized to conduct chest x-rays. There are advantages and disadvantages with regards to image quality and radiation dose when using medium and high kilovoltage (kVp) technique. However, radiographers have misconstrued understanding pertaining to the choice of tube potential as well as grid usage when performing chest radiography. Methods: The experimental study was conducted using the PBU-50 phantom by exposing it with medium kVp utilizing grid and non-grid as well as high kVp with grid. All images obtained were evaluated using the modified evaluation criteria for PA chest established by the Commission of European Communities, 1996 whilst the dose area product (DAP) was determined using the Dose Area Product (DAP) meter. The value obtained from the DAP meter was converted to entrance surface dose (ESD) using CALDOSE_X5.0 software and mathematical formula. Results: The Wilcoxon Signed-Rank Test indicated a significant difference in ESD when using medium and high kVp; Z= -2.666, p <0.05. Additionally, a significant difference in ESD was also indicated when using medium kVp with and without the grid; Z= -2.201, p <0.05. Further, Wilcoxon Signed-ranked Test indicated that there was a significant difference in image quality when using medium and high kVp; Z= -2.666, p< 0.05. Conclusions: Medium kVp should be used to demonstrate pathological conditions as well structures composed of calcium like the rib, spine or calcified lesions. Otherwise, high kVp is the preferred choice due to its low radiation dose and overall image quality acceptability.
first_indexed 2025-11-14T17:55:55Z
format Article
id iium-82902
institution International Islamic University Malaysia
institution_category Local University
language English
last_indexed 2025-11-14T17:55:55Z
publishDate 2019
publisher Kulliyyah of Allied Health Sciences, IIUM
recordtype eprints
repository_type Digital Repository
spelling iium-829022021-01-06T01:54:27Z http://irep.iium.edu.my/82902/ Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography Moey, Soo Foon Muhd Jaafar, Nur Farah Hani Saidin, Nursyahirah QC Physics Introduction: Various medium and high tube potentials were utilized to conduct chest x-rays. There are advantages and disadvantages with regards to image quality and radiation dose when using medium and high kilovoltage (kVp) technique. However, radiographers have misconstrued understanding pertaining to the choice of tube potential as well as grid usage when performing chest radiography. Methods: The experimental study was conducted using the PBU-50 phantom by exposing it with medium kVp utilizing grid and non-grid as well as high kVp with grid. All images obtained were evaluated using the modified evaluation criteria for PA chest established by the Commission of European Communities, 1996 whilst the dose area product (DAP) was determined using the Dose Area Product (DAP) meter. The value obtained from the DAP meter was converted to entrance surface dose (ESD) using CALDOSE_X5.0 software and mathematical formula. Results: The Wilcoxon Signed-Rank Test indicated a significant difference in ESD when using medium and high kVp; Z= -2.666, p <0.05. Additionally, a significant difference in ESD was also indicated when using medium kVp with and without the grid; Z= -2.201, p <0.05. Further, Wilcoxon Signed-ranked Test indicated that there was a significant difference in image quality when using medium and high kVp; Z= -2.666, p< 0.05. Conclusions: Medium kVp should be used to demonstrate pathological conditions as well structures composed of calcium like the rib, spine or calcified lesions. Otherwise, high kVp is the preferred choice due to its low radiation dose and overall image quality acceptability. Kulliyyah of Allied Health Sciences, IIUM 2019-12-31 Article PeerReviewed application/pdf en http://irep.iium.edu.my/82902/1/274-Article%20Text-1915-1-10-20200304%20%282%29%20Farah%20Hanim.pdf Moey, Soo Foon and Muhd Jaafar, Nur Farah Hani and Saidin, Nursyahirah (2019) Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography. International Journal of Allied Health Sciences (IJAHS), 3 (4). pp. 954-966. E-ISSN 2600-8491 https://journals.iium.edu.my/ijahs/index.php/IJAHS/article/view/274/350
spellingShingle QC Physics
Moey, Soo Foon
Muhd Jaafar, Nur Farah Hani
Saidin, Nursyahirah
Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography
title Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography
title_full Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography
title_fullStr Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography
title_full_unstemmed Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography
title_short Image quality and radiation dose comparison in medium and high kilovoltage for digital PA chest radiography
title_sort image quality and radiation dose comparison in medium and high kilovoltage for digital pa chest radiography
topic QC Physics
url http://irep.iium.edu.my/82902/
http://irep.iium.edu.my/82902/
http://irep.iium.edu.my/82902/1/274-Article%20Text-1915-1-10-20200304%20%282%29%20Farah%20Hanim.pdf