Accuracy of contact lens method by spherical and aspheric rigid gas permeable lenses on corneal power determination in normal eyes
Contact lens method (CLM) is an alternative option to measure corneal power by evaluating the difference of patient’s over-refraction with rigid gas permeable (RGP) lens to manifest refraction. The purpose of this study was to evaluate the accuracy of CLM using spherical (CLMspherical) and asphe...
| Main Authors: | , , , |
|---|---|
| Format: | Article |
| Language: | English English English |
| Published: |
Penerbit UKM
2020
|
| Subjects: | |
| Online Access: | http://irep.iium.edu.my/80499/ http://irep.iium.edu.my/80499/7/80499_Accuracy%20of%20Contact%20Lens%20Method%20by%20Spherical%20and%20Aspheric%20Rigid%20Gas%20Permeable_Scopus.pdf http://irep.iium.edu.my/80499/13/80499_Accuracy%20of%20Contact%20Lens%20Method%20by%20Spherical%20_wos.pdf http://irep.iium.edu.my/80499/19/80499_Accuracy%20of%20Contact%20Lens%20Method%20by%20Spherical%20and%20Aspheric.pdf |
| Summary: | Contact lens method (CLM) is an alternative option to measure corneal power by evaluating the difference of patient’s
over-refraction with rigid gas permeable (RGP) lens to manifest refraction. The purpose of this study was to evaluate
the accuracy of CLM using spherical (CLMspherical) and aspheric (CLMaspheric) RGP lenses in measuring corneal refractive
power of normal corneas. This prospective study recruited 45 normal eyes of 45 healthy subjects. The corneal
power measurements were determined by CLMspherical using Boston ES RGP and CLMaspheric using Boston Envision
RGP based on alignment fitting strategy. Manifest refraction and over-refraction were determined using a standard
procedure of objective and subjective refraction methods. IOLMaster was set as the reference method for comparison.
The mean arithmetic difference, mean absolute difference and 95% limits of agreement (LOA) of corneal powers
obtained from CLMspherical and CLMaspheric to IOLMaster value were evaluated for the accuracy assessment. The mean
arithmetic difference and mean absolute difference of corneal power s obtained from CLMspherical and CLMaspheric to
IOLMaster values were 0.10 ± 0.21 D and 0.20 ± 0.11 D, and 0.04 ± 0.09 D and 0.08 ± 0.05 D, respectively. The 95%
LOA between CLMspherical and IOLMaster ranged from -0.30 to 0.51 D, whereas between CLMaspheric and IOLMaster was
ranging from -0.14 to 0.21 D. CLM in estimating corneal power is more accurate with application of aspheric RGP
compared to spherical RGP. Hence, aspheric RGP is suggested for CLM when determining corneal power in normal
eyes. |
|---|