Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa

The value of using lead filters to improve the quality of PET data has been investigated for a CTI/Siemens 953B brain scanner operated without septa. The filters consisted of sheets of thickness 0.5mm and 1.0mm and axial width 18cm which were placed around the outer surface of the patient port. Th...

Full description

Bibliographic Details
Main Authors: Spinks, Terry Jones, Sayed, Inayatullah Shah
Format: Article
Language:English
Published: IEEE 1993
Subjects:
Online Access:http://irep.iium.edu.my/6061/
http://irep.iium.edu.my/6061/1/Lead_filters_NeuroPET.pdf
_version_ 1848776607794200576
author Spinks, Terry Jones
Sayed, Inayatullah Shah
author_facet Spinks, Terry Jones
Sayed, Inayatullah Shah
author_sort Spinks, Terry Jones
building IIUM Repository
collection Online Access
description The value of using lead filters to improve the quality of PET data has been investigated for a CTI/Siemens 953B brain scanner operated without septa. The filters consisted of sheets of thickness 0.5mm and 1.0mm and axial width 18cm which were placed around the outer surface of the patient port. The port diameter is 35 cm and the axial FOV 10.65cm. The ratios of dead times (with/without filter) with lmCi (20cm uniform cylinder) in the FOV are 0.92 (0.5mm) and 0.82 (1mm) at a lower threshold (LLD) of 380keV; the values at 15OkeV are 0.87 and 0.76. Randoms rates are reduced by 12% (0.5mm)and 27% (1mm) at a LLD of 38OkeV and by 20% and 40% at 150keV. These values reflect the significant reduction of lower energy events observed in the measured energy spectrum. However, these reductions are accompanied by a drop in the true coincidence rate such that NEC is reduced for count rates up to the maximum measured trues rate at lower thresholds of 380 keV and 150 keV. It is concluded that the filters used in this study provide no advantage in the performance of the tomograph.
first_indexed 2025-11-14T14:32:47Z
format Article
id iium-6061
institution International Islamic University Malaysia
institution_category Local University
language English
last_indexed 2025-11-14T14:32:47Z
publishDate 1993
publisher IEEE
recordtype eprints
repository_type Digital Repository
spelling iium-60612013-07-10T07:20:24Z http://irep.iium.edu.my/6061/ Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa Spinks, Terry Jones Sayed, Inayatullah Shah QC Physics The value of using lead filters to improve the quality of PET data has been investigated for a CTI/Siemens 953B brain scanner operated without septa. The filters consisted of sheets of thickness 0.5mm and 1.0mm and axial width 18cm which were placed around the outer surface of the patient port. The port diameter is 35 cm and the axial FOV 10.65cm. The ratios of dead times (with/without filter) with lmCi (20cm uniform cylinder) in the FOV are 0.92 (0.5mm) and 0.82 (1mm) at a lower threshold (LLD) of 380keV; the values at 15OkeV are 0.87 and 0.76. Randoms rates are reduced by 12% (0.5mm)and 27% (1mm) at a LLD of 38OkeV and by 20% and 40% at 150keV. These values reflect the significant reduction of lower energy events observed in the measured energy spectrum. However, these reductions are accompanied by a drop in the true coincidence rate such that NEC is reduced for count rates up to the maximum measured trues rate at lower thresholds of 380 keV and 150 keV. It is concluded that the filters used in this study provide no advantage in the performance of the tomograph. IEEE 1993-08 Article PeerReviewed application/pdf en http://irep.iium.edu.my/6061/1/Lead_filters_NeuroPET.pdf Spinks, Terry Jones and Sayed, Inayatullah Shah (1993) Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa. IEEE Transactions on Nuclear Science, 40 (4). pp. 1087-1091. ISSN 0018-9499 http://ewh.ieee.org/soc/nps/TNS.htm
spellingShingle QC Physics
Spinks, Terry Jones
Sayed, Inayatullah Shah
Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa
title Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa
title_full Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa
title_fullStr Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa
title_full_unstemmed Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa
title_short Effect of lead filters on the performance of a neuro-pet tomograph operated without Septa
title_sort effect of lead filters on the performance of a neuro-pet tomograph operated without septa
topic QC Physics
url http://irep.iium.edu.my/6061/
http://irep.iium.edu.my/6061/
http://irep.iium.edu.my/6061/1/Lead_filters_NeuroPET.pdf