A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea

© 2015, American Academy of Sleep Medicine. All rights reserved. Study Objectives: We hypothesized that a dual-channel portable monitor (PM) device could accurately identify patients who have a high pretest probability of obstructive sleep apnea (OSA), and we evaluated factors that may contribute to...

Full description

Bibliographic Details
Main Authors: Ward, Kim, McArdle, N., James, A., Bremner, A., Simpson, L., Cooper, M., Palmer, L., Fedson, A., Mukherjee, S., Hillman, D.
Format: Journal Article
Published: 2015
Online Access:http://hdl.handle.net/20.500.11937/66875
_version_ 1848761415507116032
author Ward, Kim
McArdle, N.
James, A.
Bremner, A.
Simpson, L.
Cooper, M.
Palmer, L.
Fedson, A.
Mukherjee, S.
Hillman, D.
author_facet Ward, Kim
McArdle, N.
James, A.
Bremner, A.
Simpson, L.
Cooper, M.
Palmer, L.
Fedson, A.
Mukherjee, S.
Hillman, D.
author_sort Ward, Kim
building Curtin Institutional Repository
collection Online Access
description © 2015, American Academy of Sleep Medicine. All rights reserved. Study Objectives: We hypothesized that a dual-channel portable monitor (PM) device could accurately identify patients who have a high pretest probability of obstructive sleep apnea (OSA), and we evaluated factors that may contribute to variability between PM and polysomnography (PSG) results. Methods: Consecutive clinic patients (N = 104) with possible OSA completed a home PM study, a PM study simultaneous with laboratory PSG, and a second home PM study. Uniform data analysis methods were applied to both PM and PSG data. Primary outcomes of interest were the positive likelihood ratio (LR+) and sensitivity of the PM device to "rule-in" OSA, defined as an apnea-hypopnea index (AHI) = 5 events/h on PSG. Effects of different test environment and study nights, and order of study and analysis methods (manual compared to automated) on PM diagnostic accuracy were assessed. Results: The PM has adequate LR+ (4.8), sensitivity (80%), and specificity (83%) for detecting OSA in the unattended home setting when benchmarked against laboratory PSG, with better LR+ ( > 5) and specificity (100%) and unchanged sensitivity (80%) in the simultaneous laboratory comparison. There were no significant night-night (all p > 0.10) or study order effects (home or laboratory first, p = 0.08) on AHI measures. Manual PM data review improved case finding accuracy, although this was not statistically significant (all p > 0.07). Misclassification was more frequent where OSA was mild. Conclusions: Overall performance of the PM device is consistent with current recommended criteria for an "acceptable" device to confidently "rule-in" OSA (AHI = 5 events/h) in a high pretest probability clinic population. Our data support the utility of simple two-channel diagnostic devices to confirm the diagnosis of OSA in the home environment.
first_indexed 2025-11-14T10:31:19Z
format Journal Article
id curtin-20.500.11937-66875
institution Curtin University Malaysia
institution_category Local University
last_indexed 2025-11-14T10:31:19Z
publishDate 2015
recordtype eprints
repository_type Digital Repository
spelling curtin-20.500.11937-668752018-05-18T08:07:10Z A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea Ward, Kim McArdle, N. James, A. Bremner, A. Simpson, L. Cooper, M. Palmer, L. Fedson, A. Mukherjee, S. Hillman, D. © 2015, American Academy of Sleep Medicine. All rights reserved. Study Objectives: We hypothesized that a dual-channel portable monitor (PM) device could accurately identify patients who have a high pretest probability of obstructive sleep apnea (OSA), and we evaluated factors that may contribute to variability between PM and polysomnography (PSG) results. Methods: Consecutive clinic patients (N = 104) with possible OSA completed a home PM study, a PM study simultaneous with laboratory PSG, and a second home PM study. Uniform data analysis methods were applied to both PM and PSG data. Primary outcomes of interest were the positive likelihood ratio (LR+) and sensitivity of the PM device to "rule-in" OSA, defined as an apnea-hypopnea index (AHI) = 5 events/h on PSG. Effects of different test environment and study nights, and order of study and analysis methods (manual compared to automated) on PM diagnostic accuracy were assessed. Results: The PM has adequate LR+ (4.8), sensitivity (80%), and specificity (83%) for detecting OSA in the unattended home setting when benchmarked against laboratory PSG, with better LR+ ( > 5) and specificity (100%) and unchanged sensitivity (80%) in the simultaneous laboratory comparison. There were no significant night-night (all p > 0.10) or study order effects (home or laboratory first, p = 0.08) on AHI measures. Manual PM data review improved case finding accuracy, although this was not statistically significant (all p > 0.07). Misclassification was more frequent where OSA was mild. Conclusions: Overall performance of the PM device is consistent with current recommended criteria for an "acceptable" device to confidently "rule-in" OSA (AHI = 5 events/h) in a high pretest probability clinic population. Our data support the utility of simple two-channel diagnostic devices to confirm the diagnosis of OSA in the home environment. 2015 Journal Article http://hdl.handle.net/20.500.11937/66875 10.5664/jcsm.4600 restricted
spellingShingle Ward, Kim
McArdle, N.
James, A.
Bremner, A.
Simpson, L.
Cooper, M.
Palmer, L.
Fedson, A.
Mukherjee, S.
Hillman, D.
A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea
title A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea
title_full A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea
title_fullStr A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea
title_full_unstemmed A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea
title_short A comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea
title_sort comprehensive evaluation of a two-channel portable monitor to "rule in" obstructive sleep apnea
url http://hdl.handle.net/20.500.11937/66875