Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study

Objective: To identify the optimal dual-energy computed tomography (DECT) scanning protocol for peripheral arterial stents while achieving a low radiation dose, while still maintaining diagnostic image quality, as determined by an in vitro phantom study. Methods: Dual-energy scans in monochromatic s...

Full description

Bibliographic Details
Main Authors: Almutairi, Abdulrahman, Sun, Zhonghua, Safran, Z., Poovathumkadavi, A., Albader, S., Ifdailat, H.
Format: Journal Article
Published: MDPI AG 2015
Online Access:http://hdl.handle.net/20.500.11937/63189
_version_ 1848761018121977856
author Almutairi, Abdulrahman
Sun, Zhonghua
Safran, Z.
Poovathumkadavi, A.
Albader, S.
Ifdailat, H.
author_facet Almutairi, Abdulrahman
Sun, Zhonghua
Safran, Z.
Poovathumkadavi, A.
Albader, S.
Ifdailat, H.
author_sort Almutairi, Abdulrahman
building Curtin Institutional Repository
collection Online Access
description Objective: To identify the optimal dual-energy computed tomography (DECT) scanning protocol for peripheral arterial stents while achieving a low radiation dose, while still maintaining diagnostic image quality, as determined by an in vitro phantom study. Methods: Dual-energy scans in monochromatic spectral imaging mode were performed on a peripheral arterial phantom with use of three gemstone spectral imaging (GSI) protocols, three pitch values, and four kiloelectron volts (keV) ranges. A total of 15 stents of different sizes, materials, and designs were deployed in the phantom. Image noise, the signal-to-noise ratio (SNR), different levels of adaptive statistical iterative reconstruction (ASIR), and the four levels of monochromatic energy for DECT imaging of peripheral arterial stents were measured and compared to determine the optimal protocols. Results: A total of 36 scans with 180 datasets were reconstructed from a combination of different protocols. There was a significant reduction of image noise with a higher SNR from monochromatic energy images between 65 and 70 keV in all investigated preset GSI protocols (p < 0.05). In addition, significant effects were found from the main effect analysis for these factors: GSI, pitch, and keV (p = 0.001). In contrast, there was significant interaction on the unstented area between GSI and ASIR (p = 0.015) and a very high significant difference between keV and ASIR (p < 0.001). A radiation dose reduction of 50% was achieved. Conclusions: The optimal scanning protocol and energy level in the phantom study were GSI-48, pitch value 0.984, and 65 keV, which resulted in lower image noise and a lower radiation dose, but with acceptable diagnostic images.
first_indexed 2025-11-14T10:25:00Z
format Journal Article
id curtin-20.500.11937-63189
institution Curtin University Malaysia
institution_category Local University
last_indexed 2025-11-14T10:25:00Z
publishDate 2015
publisher MDPI AG
recordtype eprints
repository_type Digital Repository
spelling curtin-20.500.11937-631892018-04-04T07:09:42Z Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study Almutairi, Abdulrahman Sun, Zhonghua Safran, Z. Poovathumkadavi, A. Albader, S. Ifdailat, H. Objective: To identify the optimal dual-energy computed tomography (DECT) scanning protocol for peripheral arterial stents while achieving a low radiation dose, while still maintaining diagnostic image quality, as determined by an in vitro phantom study. Methods: Dual-energy scans in monochromatic spectral imaging mode were performed on a peripheral arterial phantom with use of three gemstone spectral imaging (GSI) protocols, three pitch values, and four kiloelectron volts (keV) ranges. A total of 15 stents of different sizes, materials, and designs were deployed in the phantom. Image noise, the signal-to-noise ratio (SNR), different levels of adaptive statistical iterative reconstruction (ASIR), and the four levels of monochromatic energy for DECT imaging of peripheral arterial stents were measured and compared to determine the optimal protocols. Results: A total of 36 scans with 180 datasets were reconstructed from a combination of different protocols. There was a significant reduction of image noise with a higher SNR from monochromatic energy images between 65 and 70 keV in all investigated preset GSI protocols (p < 0.05). In addition, significant effects were found from the main effect analysis for these factors: GSI, pitch, and keV (p = 0.001). In contrast, there was significant interaction on the unstented area between GSI and ASIR (p = 0.015) and a very high significant difference between keV and ASIR (p < 0.001). A radiation dose reduction of 50% was achieved. Conclusions: The optimal scanning protocol and energy level in the phantom study were GSI-48, pitch value 0.984, and 65 keV, which resulted in lower image noise and a lower radiation dose, but with acceptable diagnostic images. 2015 Journal Article http://hdl.handle.net/20.500.11937/63189 10.3390/ijms160511531 http://creativecommons.org/licenses/by/4.0/ MDPI AG fulltext
spellingShingle Almutairi, Abdulrahman
Sun, Zhonghua
Safran, Z.
Poovathumkadavi, A.
Albader, S.
Ifdailat, H.
Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study
title Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study
title_full Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study
title_fullStr Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study
title_full_unstemmed Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study
title_short Optimal Scanning Protocols for Dual-Energy CT Angiography in Peripheral Arterial Stents: An in Vitro Phantom Study
title_sort optimal scanning protocols for dual-energy ct angiography in peripheral arterial stents: an in vitro phantom study
url http://hdl.handle.net/20.500.11937/63189