Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires

Purpose: To compare in vivo the use of synchrotron radiation to computed tomography angiography (CTA) for the measurement of cross-sectional area (CSA) reduction of the aortic branch ostia caused by suprarenal stent-graft wires. Methods: This study was performed with a Zenith stent-graft placed in a...

Full description

Bibliographic Details
Main Authors: Sun, Zhonghua, Ng, Curtise
Format: Journal Article
Published: International Society of Endovascular Specialists 2017
Online Access:http://hdl.handle.net/20.500.11937/58915
_version_ 1848760375467573248
author Sun, Zhonghua
Ng, Curtise
author_facet Sun, Zhonghua
Ng, Curtise
author_sort Sun, Zhonghua
building Curtin Institutional Repository
collection Online Access
description Purpose: To compare in vivo the use of synchrotron radiation to computed tomography angiography (CTA) for the measurement of cross-sectional area (CSA) reduction of the aortic branch ostia caused by suprarenal stent-graft wires. Methods: This study was performed with a Zenith stent-graft placed in a phantom of the human aorta to simulate treatment of abdominal aortic aneurysm. Synchrotron radiation scans were performed using beam energies between 40 and 100 keV and spatial resolution of 19.88 µm per pixel. CSA reduction of the aortic branch ostia by suprarenal stent wires was calculated based on these exposure factors and compared with measurements from CTA images acquired on a 64-row scanner with slice thicknesses of 1.0, 1.5, and 2.0 mm. Results: Images acquired with synchrotron radiation showed <10% of the CSA occupied by stent wires when a single wire crossed a renal artery ostium and <20% for 2 wires crossing a renovisceral branch ostium. The corresponding areas ranged from 24% to 25% for a single wire and from 40% to 48% for double wires crossing the branch ostia when measured on CT images. The stent wire was accurately assessed on synchrotron radiation with a diameter between 0.38±0.01 and 0.53±0.03 mm, which is close to the actual size of 0.47±0.01 mm. The wire diameter measured on CT images was greatly overestimated (1.15±0.01 to 1.57±0.02 mm). Conclusion: CTA has inferior spatial resolution that hinders accurate assessment of CSA reduction. This experiment demonstrated the superiority of synchrotron radiation over CTA for more accurate assessment of aortic stent wires and CSA reduction of the aortic branch ostia.
first_indexed 2025-11-14T10:14:47Z
format Journal Article
id curtin-20.500.11937-58915
institution Curtin University Malaysia
institution_category Local University
last_indexed 2025-11-14T10:14:47Z
publishDate 2017
publisher International Society of Endovascular Specialists
recordtype eprints
repository_type Digital Repository
spelling curtin-20.500.11937-589152018-03-20T04:35:00Z Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires Sun, Zhonghua Ng, Curtise Purpose: To compare in vivo the use of synchrotron radiation to computed tomography angiography (CTA) for the measurement of cross-sectional area (CSA) reduction of the aortic branch ostia caused by suprarenal stent-graft wires. Methods: This study was performed with a Zenith stent-graft placed in a phantom of the human aorta to simulate treatment of abdominal aortic aneurysm. Synchrotron radiation scans were performed using beam energies between 40 and 100 keV and spatial resolution of 19.88 µm per pixel. CSA reduction of the aortic branch ostia by suprarenal stent wires was calculated based on these exposure factors and compared with measurements from CTA images acquired on a 64-row scanner with slice thicknesses of 1.0, 1.5, and 2.0 mm. Results: Images acquired with synchrotron radiation showed <10% of the CSA occupied by stent wires when a single wire crossed a renal artery ostium and <20% for 2 wires crossing a renovisceral branch ostium. The corresponding areas ranged from 24% to 25% for a single wire and from 40% to 48% for double wires crossing the branch ostia when measured on CT images. The stent wire was accurately assessed on synchrotron radiation with a diameter between 0.38±0.01 and 0.53±0.03 mm, which is close to the actual size of 0.47±0.01 mm. The wire diameter measured on CT images was greatly overestimated (1.15±0.01 to 1.57±0.02 mm). Conclusion: CTA has inferior spatial resolution that hinders accurate assessment of CSA reduction. This experiment demonstrated the superiority of synchrotron radiation over CTA for more accurate assessment of aortic stent wires and CSA reduction of the aortic branch ostia. 2017 Journal Article http://hdl.handle.net/20.500.11937/58915 10.1177/1526602817732315 International Society of Endovascular Specialists fulltext
spellingShingle Sun, Zhonghua
Ng, Curtise
Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires
title Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires
title_full Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires
title_fullStr Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires
title_full_unstemmed Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires
title_short Use of Synchrotron Radiation to Accurately Assess Cross-Sectional Area Reduction of the Aortic Branch Ostia Caused by Suprarenal Stent Wires
title_sort use of synchrotron radiation to accurately assess cross-sectional area reduction of the aortic branch ostia caused by suprarenal stent wires
url http://hdl.handle.net/20.500.11937/58915