A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy

Background: The aim was to assess whether inserting a longer soft silicone short-term dialysis catheter targeting tip placement in the right atrium could improve dialyzer circuit life span compared with inserting a shorter dialysis catheter targeting tip placement in the superior vena cava. Study De...

Full description

Bibliographic Details
Main Authors: Morgan, D., Ho, K., Murray, C., Davies, Hugh, Louw, J.
Format: Journal Article
Published: 2012
Online Access:http://hdl.handle.net/20.500.11937/31050
_version_ 1848753266767167488
author Morgan, D.
Ho, K.
Murray, C.
Davies, Hugh
Louw, J.
author_facet Morgan, D.
Ho, K.
Murray, C.
Davies, Hugh
Louw, J.
author_sort Morgan, D.
building Curtin Institutional Repository
collection Online Access
description Background: The aim was to assess whether inserting a longer soft silicone short-term dialysis catheter targeting tip placement in the right atrium could improve dialyzer circuit life span compared with inserting a shorter dialysis catheter targeting tip placement in the superior vena cava. Study Design: Randomized unblinded controlled study. Setting & Participants: A tertiary multidisciplinary intensive care unit enrolling 100 critically ill patients requiring continuous renal replacement therapy (CRRT). Intervention: Placement of longer (20-24 cm) versus shorter dialysis catheters (15-20 cm) within one of the major thoracic veins for initiation of CRRT. Outcomes: The primary study outcome was duration of dialysis circuit life span. Secondary outcomes included delivered daily dialysis dose, incidence and cause of CRRT circuit failure, complications potentially related to the position of the short-term dialysis catheter, mortality, and patient length of stay.Results: Placing the longer dialysis catheters was associated with an increased average dialyzer life span of 6.5 hours (24 hours [25th-75th percentile, 11-32] vs 17.5 hours [25th-75th percentile, 8-23]; P = 0.001), improved delivered daily dialysis dose (91% [25th-75th percentile, 85%-100%] vs 81% [25th-75th percentile, 72%-97%]; < 0.001), and reduced number of dialyzers clotted (2.3 vs 3.6; P = 0.04) or circuits taken down due to vascular access problem (0.19 vs 0.53; P = 0.04) per patient compared with placing shorter dialysis catheters. The incidence of atrial arrhythmias was similar between groups (28% vs 21%; P = 0.6) and the only mechanical complication was the malposition of one dialysis catheter tip in the longer dialysis catheter group. Limitations: Single-center study design. Conclusions: The use of longer soft silicone short-term dialysis catheters targeting right atrial placement appeared to be safe and could improve dialyzer life span and daily dialysis dose of CRRT delivered compared with the use of shorter catheters targeting superior vena cava placement.
first_indexed 2025-11-14T08:21:47Z
format Journal Article
id curtin-20.500.11937-31050
institution Curtin University Malaysia
institution_category Local University
last_indexed 2025-11-14T08:21:47Z
publishDate 2012
recordtype eprints
repository_type Digital Repository
spelling curtin-20.500.11937-310502017-09-13T15:11:18Z A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy Morgan, D. Ho, K. Murray, C. Davies, Hugh Louw, J. Background: The aim was to assess whether inserting a longer soft silicone short-term dialysis catheter targeting tip placement in the right atrium could improve dialyzer circuit life span compared with inserting a shorter dialysis catheter targeting tip placement in the superior vena cava. Study Design: Randomized unblinded controlled study. Setting & Participants: A tertiary multidisciplinary intensive care unit enrolling 100 critically ill patients requiring continuous renal replacement therapy (CRRT). Intervention: Placement of longer (20-24 cm) versus shorter dialysis catheters (15-20 cm) within one of the major thoracic veins for initiation of CRRT. Outcomes: The primary study outcome was duration of dialysis circuit life span. Secondary outcomes included delivered daily dialysis dose, incidence and cause of CRRT circuit failure, complications potentially related to the position of the short-term dialysis catheter, mortality, and patient length of stay.Results: Placing the longer dialysis catheters was associated with an increased average dialyzer life span of 6.5 hours (24 hours [25th-75th percentile, 11-32] vs 17.5 hours [25th-75th percentile, 8-23]; P = 0.001), improved delivered daily dialysis dose (91% [25th-75th percentile, 85%-100%] vs 81% [25th-75th percentile, 72%-97%]; < 0.001), and reduced number of dialyzers clotted (2.3 vs 3.6; P = 0.04) or circuits taken down due to vascular access problem (0.19 vs 0.53; P = 0.04) per patient compared with placing shorter dialysis catheters. The incidence of atrial arrhythmias was similar between groups (28% vs 21%; P = 0.6) and the only mechanical complication was the malposition of one dialysis catheter tip in the longer dialysis catheter group. Limitations: Single-center study design. Conclusions: The use of longer soft silicone short-term dialysis catheters targeting right atrial placement appeared to be safe and could improve dialyzer life span and daily dialysis dose of CRRT delivered compared with the use of shorter catheters targeting superior vena cava placement. 2012 Journal Article http://hdl.handle.net/20.500.11937/31050 10.1053/j.ajkd.2012.01.021 restricted
spellingShingle Morgan, D.
Ho, K.
Murray, C.
Davies, Hugh
Louw, J.
A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy
title A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy
title_full A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy
title_fullStr A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy
title_full_unstemmed A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy
title_short A randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy
title_sort randomized trial of catheters of different lengths to achieve right atrium versus superior vena cava placement for continuous renal replacement therapy
url http://hdl.handle.net/20.500.11937/31050