Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population

Objective. To examine the intratester reliability of the Spin-T goniometer, a cervical range of motion device, in a normal Indian population. Methods. Subjects comprised 30 healthy adults with mean age of 34 years (range, 18-65 years). The subjects were stabilized in the sitting position and the Spi...

Full description

Bibliographic Details
Main Authors: Agarwal, S., Allison, Garry, Singer, K.
Format: Journal Article
Published: Elsevier Inc 2005
Subjects:
Online Access:http://hdl.handle.net/20.500.11937/23663
_version_ 1848751212801818624
author Agarwal, S.
Allison, Garry
Singer, K.
author_facet Agarwal, S.
Allison, Garry
Singer, K.
author_sort Agarwal, S.
building Curtin Institutional Repository
collection Online Access
description Objective. To examine the intratester reliability of the Spin-T goniometer, a cervical range of motion device, in a normal Indian population. Methods. Subjects comprised 30 healthy adults with mean age of 34 years (range, 18-65 years). The subjects were stabilized in the sitting position and the Spin-T goniometer mounted on the head of the subject. The study design was a within-subject repeated intratester reliability trial conducted for cervical range of motion in 6 directions of movement. Three measurements were taken in each direction (flexion, extension lateral flexion, and lateral rotation) per participant. Reliability coefficients, intraclass correlation coefficients, and 95% confidence interval were derived from repeated-measures analysis of variance (ANOVA). Where differences in ANOVA were detected, a paired t test was conducted and the typical error values and coefficient of variance were calculated. Results. All repeated measures showed high intraclass correlation coefficients (all >0.96, P <.01). The ANOVA detected no differences between trials for all movements except rotation. The typical error values for the rotation trials did not exceed 2.5 and the coefficient of variance did not exceed 4%, which is clinically acceptable considering the normally variable cervical range of movement. Conclusion. In this study, the Spin-T goniometer proved to be a reliable measuring instrument for cervical range of movement in an Indian population. The use of a laser pointer fixed to the instrument ensured a consistent neutral start position.
first_indexed 2025-11-14T07:49:09Z
format Journal Article
id curtin-20.500.11937-23663
institution Curtin University Malaysia
institution_category Local University
last_indexed 2025-11-14T07:49:09Z
publishDate 2005
publisher Elsevier Inc
recordtype eprints
repository_type Digital Repository
spelling curtin-20.500.11937-236632017-09-13T16:00:44Z Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population Agarwal, S. Allison, Garry Singer, K. Articular Reproducibility of Results Validity Cervical Vertebrae Range of Motion Objective. To examine the intratester reliability of the Spin-T goniometer, a cervical range of motion device, in a normal Indian population. Methods. Subjects comprised 30 healthy adults with mean age of 34 years (range, 18-65 years). The subjects were stabilized in the sitting position and the Spin-T goniometer mounted on the head of the subject. The study design was a within-subject repeated intratester reliability trial conducted for cervical range of motion in 6 directions of movement. Three measurements were taken in each direction (flexion, extension lateral flexion, and lateral rotation) per participant. Reliability coefficients, intraclass correlation coefficients, and 95% confidence interval were derived from repeated-measures analysis of variance (ANOVA). Where differences in ANOVA were detected, a paired t test was conducted and the typical error values and coefficient of variance were calculated. Results. All repeated measures showed high intraclass correlation coefficients (all >0.96, P <.01). The ANOVA detected no differences between trials for all movements except rotation. The typical error values for the rotation trials did not exceed 2.5 and the coefficient of variance did not exceed 4%, which is clinically acceptable considering the normally variable cervical range of movement. Conclusion. In this study, the Spin-T goniometer proved to be a reliable measuring instrument for cervical range of movement in an Indian population. The use of a laser pointer fixed to the instrument ensured a consistent neutral start position. 2005 Journal Article http://hdl.handle.net/20.500.11937/23663 10.1016/j.jmpt.2005.07.005 Elsevier Inc restricted
spellingShingle Articular
Reproducibility of Results
Validity
Cervical Vertebrae
Range of Motion
Agarwal, S.
Allison, Garry
Singer, K.
Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population
title Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population
title_full Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population
title_fullStr Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population
title_full_unstemmed Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population
title_short Reliability of the Spin-T Cervical Goniometer in Measuring Cervical Range of Motion in an Asymptomatic Indian Population
title_sort reliability of the spin-t cervical goniometer in measuring cervical range of motion in an asymptomatic indian population
topic Articular
Reproducibility of Results
Validity
Cervical Vertebrae
Range of Motion
url http://hdl.handle.net/20.500.11937/23663