Stress analysis of a centrally fractured rib fixated by an intramedullary screw

The stress on an intramedullary screw rib fixation device holding together a centrally fractured human rib under in vivo force loadings was studied using finite element analysis (FEA). Validation of the FEA modelling using pullout from porcine ribs proved FEA to be suitable for assessing the structu...

Full description

Bibliographic Details
Main Authors: Liovic, P., Sutalo, Ilija, Marasco, S.
Format: Journal Article
Published: Taylor & Francis 2012
Subjects:
Online Access:http://hdl.handle.net/20.500.11937/19581
_version_ 1848750073278627840
author Liovic, P.
Sutalo, Ilija
Marasco, S.
author_facet Liovic, P.
Sutalo, Ilija
Marasco, S.
author_sort Liovic, P.
building Curtin Institutional Repository
collection Online Access
description The stress on an intramedullary screw rib fixation device holding together a centrally fractured human rib under in vivo force loadings was studied using finite element analysis (FEA). Validation of the FEA modelling using pullout from porcine ribs proved FEA to be suitable for assessing the structural integrity of screw/bone systems such as rib fixated by a screw. In the human rib fixation investigation, it was found that intramedullary bioresorbable Bioretec screws can fixate centrally fractured human ribs under normal breathing conditions. However, under coughing conditions, simulation showed Bioretec fixating screws to bend substantially. High stresses in the screw are mainly the result of flexion induced by the force loading, and are restricted to thin regions on the outside of the screw shaft. Stiffer screws result in less locally intense stress concentrations in bone, indicating that bone failure in the bone/screw contact regions can be averted with improvements in screw stiffness.
first_indexed 2025-11-14T07:31:02Z
format Journal Article
id curtin-20.500.11937-19581
institution Curtin University Malaysia
institution_category Local University
last_indexed 2025-11-14T07:31:02Z
publishDate 2012
publisher Taylor & Francis
recordtype eprints
repository_type Digital Repository
spelling curtin-20.500.11937-195812017-09-13T16:09:10Z Stress analysis of a centrally fractured rib fixated by an intramedullary screw Liovic, P. Sutalo, Ilija Marasco, S. finite element analysis flail chest bioresorbable screw intramedullary rib fixation The stress on an intramedullary screw rib fixation device holding together a centrally fractured human rib under in vivo force loadings was studied using finite element analysis (FEA). Validation of the FEA modelling using pullout from porcine ribs proved FEA to be suitable for assessing the structural integrity of screw/bone systems such as rib fixated by a screw. In the human rib fixation investigation, it was found that intramedullary bioresorbable Bioretec screws can fixate centrally fractured human ribs under normal breathing conditions. However, under coughing conditions, simulation showed Bioretec fixating screws to bend substantially. High stresses in the screw are mainly the result of flexion induced by the force loading, and are restricted to thin regions on the outside of the screw shaft. Stiffer screws result in less locally intense stress concentrations in bone, indicating that bone failure in the bone/screw contact regions can be averted with improvements in screw stiffness. 2012 Journal Article http://hdl.handle.net/20.500.11937/19581 10.1080/10255842.2012.727402 Taylor & Francis restricted
spellingShingle finite element analysis
flail chest
bioresorbable screw
intramedullary rib fixation
Liovic, P.
Sutalo, Ilija
Marasco, S.
Stress analysis of a centrally fractured rib fixated by an intramedullary screw
title Stress analysis of a centrally fractured rib fixated by an intramedullary screw
title_full Stress analysis of a centrally fractured rib fixated by an intramedullary screw
title_fullStr Stress analysis of a centrally fractured rib fixated by an intramedullary screw
title_full_unstemmed Stress analysis of a centrally fractured rib fixated by an intramedullary screw
title_short Stress analysis of a centrally fractured rib fixated by an intramedullary screw
title_sort stress analysis of a centrally fractured rib fixated by an intramedullary screw
topic finite element analysis
flail chest
bioresorbable screw
intramedullary rib fixation
url http://hdl.handle.net/20.500.11937/19581